  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tustin",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87322",
      "recalling_firm": "Canon Medical System, USA, INC.",
      "address_1": "2441 Michelle Dr",
      "address_2": "N/A",
      "postal_code": "92780-7047",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, WA, WI and WV. The countries of  Canada, China, Japan, Korea, and Singapore.",
      "recall_number": "Z-1271-2021",
      "product_description": "Canon INTERVENTIONAL ANGIOGRAPHY SYSTEM MODEL INFX-8000F    Catheterization Table CAT-880B  Canon CATHETERIZATION TABLE, MODEL CAT-880B INPUT 220V - Product Usage: indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen, and lower extremities.",
      "product_quantity": "Total of all Systems = 145 systems (U.S.)",
      "reason_for_recall": "System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.",
      "recall_initiation_date": "20201208",
      "center_classification_date": "20210317",
      "report_date": "20210324",
      "code_info": "Table CAT-880B  Models:  INFX-8000F/AH  INFX-8000F/AH  INFX-8000F/AH  INFX-8000F/AH  INFX-8000F/AH  INFX-8000F/BHUP  INFX-8000F/AH  INFX-8000F/AH  INFX-8000F/UU  INFX-8000F/AH  INFX-8000F/AH  INFX-8000F/AH  INFX-8000F/AH  INFX-8000F/AH  INFX-8000F/AH  INFX-8000F/BH    Table Serial Numbers:  99E1582258  99F18Y2428  99C1482200  99E15Y2271  99F1862406  99G19Z2438  99E1662308  99E1762365  99A1337005  99F1852401  99E1612286  99E1712347  99E1712345  99C14Z2224  99E1592262  99G1972436    System Serial Numbers:  AHA1582003  AHC18Y2013  AHA1482001  AHA15Y2005  AHC1862012  BHB19Z2002  AHB1662007  AHB1762010  UUB12X2034  AHC1852011  AHA1612006  AHB1712009  AHB1712008  AHA14Z2002  AHA1592004  BHB1972001"
    }
  ]
}