  "results": [
    {
      "status": "Terminated",
      "city": "EDEN PRAIRIE",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73468",
      "recalling_firm": "Laserex Systems Inc.",
      "address_1": "7138 SHADY OAK RD",
      "address_2": "N/A",
      "postal_code": "55344-3517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Distribution",
      "recall_number": "Z-1270-2016",
      "product_description": "Ellex Medical Fixed eye Safety filters Model  Integre S LP5532.    The filter is located in the binocular mount of the laser delivery head. The filter prevents transmission of treatment  laser light to physician eyes while the  aiming beam to remain visible.      .",
      "product_quantity": "72 units plus 2 spare units",
      "reason_for_recall": "The Firm discovered that there is a design error which is traceable back to 2012 when a design change on the safety filter optic adaptor was implemented.",
      "recall_initiation_date": "20160201",
      "center_classification_date": "20160407",
      "termination_date": "20161027",
      "report_date": "20160413",
      "code_info": "Model Number : 3143270",
      "more_code_info": ""
    }
  ]
}