  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67607",
      "recalling_firm": "Lin-Zhi International Inc",
      "address_1": "670 Almanor Ave",
      "address_2": "N/A",
      "postal_code": "94085-3513",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Distribution including CA.",
      "recall_number": "Z-1270-2014",
      "product_description": "Single Analyte DAU Calibrator (5 mL)  Norbuprenorphine: Cutoff Calibrator;  IVD - for in vitro diagnostic use only;  Lin-Zhi International  Sunnyvale, CA    The Norbuprenorphine Drug of Abuse (DAU) cutoff Calibrator is used for qualitative and semi-qualitative calibration of the Lin-Zhi Internal, Inc. Buprenorphine (BUP) enzyme Immunoassay (REF # A53684) on the Synchron Clinical Analyzer",
      "product_quantity": "120 units, 60 of each lot number",
      "reason_for_recall": "Customer reported that the Synchron NBUP cutoff calibrator (10 ng/mL) could not be qualified through quality control testing. The cutoff calibrator was recovering at a concentration close to the NBUP Control Level 1 (7 ng/mL).",
      "recall_initiation_date": "20140218",
      "center_classification_date": "20140321",
      "termination_date": "20140425",
      "report_date": "20140402",
      "code_info": "Product catalog number: A68827;  Lots 1308006, 1308131, Expiry: March 28, 2015."
    }
  ]
}