  "results": [
    {
      "country": "United States",
      "city": "Bedford",
      "address_1": "5 Oak Park Dr",
      "reason_for_recall": "Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.",
      "address_2": "",
      "product_quantity": "",
      "code_info": "Medica s Lot # 16292",
      "center_classification_date": "20180328",
      "distribution_pattern": "USA: MI  FL  CA  TX    India, Russia, Puerto Rico, N. Mariana Island/Palau.",
      "state": "MA",
      "product_description": "EasyCAL HbA1c Calibrator; REF 10658  Kit is packaged in a paper box containing 4 vials of 0.25ml of calibrant, one for each calibration level, and an insert sheet.",
      "report_date": "20180404",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medica Corporation",
      "recall_number": "Z-1267-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "77160",
      "termination_date": "20190820",
      "more_code_info": "",
      "recall_initiation_date": "20170411",
      "postal_code": "01730-1430",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}