  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82582",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK",
      "recall_number": "Z-1264-2019",
      "product_description": "NM/CT 870 CZT, Model/Catalogue/Code # H3906CW    Product Usage:  The GE Discovery NM/CT 670 CZT system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and anatomical localization of SPECT images and for standalone head, whole body, cardiac and vascular X-ray Computed Tomography applications.",
      "product_quantity": "16 units",
      "reason_for_recall": "During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.",
      "recall_initiation_date": "20190320",
      "center_classification_date": "20190502",
      "termination_date": "20201005",
      "report_date": "20190508",
      "code_info": "Mfg. Lot or Serial # System ID  87CY63017 SHIPPED (AO 1630621)  87CX63007 AH0101NM05  87CX63014 030078NU28  87CX63008 190020NU01  87CX63006 030240NU11  87CX63015 705743NM870  87CX63011 M4184879  87CX63003 M9886873  87CX63004 X171582003  87CX63009 M171582018  87CX63010 UC1796NU01  87CX63013 NO1068MN18  87CX63005 A004NU20  87CX63002 C028NU01  87CX63012 C028NU02  87CX63001 900010NU12"
    }
  ]
}