  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Avondale",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73543",
      "recalling_firm": "Wfr/aquaplast Corp",
      "address_1": "440 Church Rd",
      "address_2": "N/A",
      "postal_code": "19311-9786",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of NY, PA, CA, IL, and TX.    Foreign Countries: Singapore, Australia, Saudi Arabia, and Spain.",
      "recall_number": "Z-1264-2016",
      "product_description": "Clear Vue Prone Breast Treatment Table and Access ClearVision Prone Breast Device     Product Usage:  Clear Vue Prone Breast Treatment Table and Access ClearVision Prone Breast Device is a medical charged-particle radiation therapy system. A breast radiation therapy patient support, is an accessory for medical linear accelerator radiation therapy. It may be used with CT and MR scanners to acquire images for radiation therapy planning.",
      "product_quantity": "18 units of RT-4545CV-17 and 12 units of RT-4545CV-20",
      "reason_for_recall": "Some of the Access ClearVue Prone Breast devices may contain scales that are misaligned from the zero reference with respect to other Access ClearVue Prone Breast devices.  This can result in differences in scale orientation.",
      "recall_initiation_date": "20160107",
      "center_classification_date": "20160328",
      "termination_date": "20170106",
      "report_date": "20160406",
      "code_info": "device model numbers: RT-4545CV-17, RT-4545CV-20"
    }
  ]
}