  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90195",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "CA FL MA MD MI NJ OH PR WI",
      "recall_number": "Z-1263-2022",
      "product_description": "Olympus URETERO-RENO FIBERSCOPE  Model: URF-P6",
      "product_quantity": "2 units",
      "reason_for_recall": "Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.",
      "recall_initiation_date": "20220420",
      "center_classification_date": "20220613",
      "report_date": "20220622",
      "code_info": "UDI-DI: 04953170340802 Serial Number: 2836210 2834747I",
      "more_code_info": ""
    }
  ]
}