  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82582",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK",
      "recall_number": "Z-1263-2019",
      "product_description": "NM/CT 860      Product Usage:    The GE NM/CT 860 system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and anatomical localization of SPECT images and for standalone head, whole body and vascular X-ray Computed Tomography applications.",
      "product_quantity": "6 units",
      "reason_for_recall": "During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.",
      "recall_initiation_date": "20190320",
      "center_classification_date": "20190502",
      "termination_date": "20201005",
      "report_date": "20190508",
      "code_info": "Mfg. Lot or Serial # System ID Model, Catalogue or Code #  860Y62005 SHIPPED (GON 4736850) H3908AD  850X61003 401793860NM H3908AD  860X62001 E6843969 H3908AE  870X64005 HC4387NU05 H3100AS  860Y62004 083026212959219   H3908AD  860X62002 PL0399NM01 H3908AD"
    }
  ]
}