      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70264",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) and the countries of  Canada, Australia and England.",
      "recall_number": "Z-1263-2015",
      "product_description": "VITROS Hand-held Barcode Scanner, Catalog Number 6844210, for use with the following VITROS Systems: VITROS 5,1 FS Chemistry System, VITROS 3600 Immunodiagnostics System, VITROS 4600 Chemistry System, VITROS 5600 Integrated System.  This product is an accessory for use with VITROS Analyzer Systems.",
      "product_quantity": "Domestic: 62 units; Foreign: 10 units",
      "reason_for_recall": "In combination with the Batch Programming option (only) for programming samples, the hand-held barcode scanner unexpectedly skips sample cup positions within a sample tray. This issue only occurs when using the Batch Programming option; all other sample programming options function as intended.",
      "recall_initiation_date": "20141229",
      "center_classification_date": "20150311",
      "termination_date": "20170327",
      "report_date": "20150318",
      "code_info": "Lot Numbers: 0914F11968, 0914F11969, 0914F11970, 0914F11971, 0914F11972, 0914F11973, 0914F11974, 0914F11975, 0914F11976, 0914F11977, 0914F11978, 0914F11979, 0914F11980, 0914F11981, 0914F11982, 0914F11983, 0914F11984, 0914F11985, 0914F11986,  0914F11987, 0914F11988, 0914F11989, 0914F11990, 0914F11997, 0914F12007, 0914F12016, 0914F12017, 0914F12020, 0914F12022, 0914F12028, 0914F12033, 0914F12037, 0914F12039, 0914F12040, 0914F12041, 0914F12046, 0914F12048, 0914F12049, 0914F12050,  0914F12053, 0914F12054, 0914F12055, 0914F12056, 0914F12057, 0914F12058, 0914F12059, 0914F12060, 0914F12061, 0914F12062, 0914F12063, 0914F12064, 0914F12065, 0914F12066, 0914F12067, 0914F12081, 0914F12082, 0914F12083, 0914F12084, 0914F12085, 0914F12086, 1014F12318, 1014F12319, 1014F12320, 1014F12321, 1014F12322, 1014F12323, 1014F12386, 1014F12387, 1014F12388, 1014F12389, 1014F12390, 1014F12391."
    }
  ]
}