  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91505",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 101",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of FL, GA, WA and the countries of  United Kingdom, Germany, France",
      "recall_number": "Z-1262-2023",
      "product_description": "Patient Side Cart, REF: 380620, with instrument arms, used as part of the da Vinci X Surgical System, REF: IS4200;    Patient Side Cart, REF: 380652, with instrument arms, used as part of the da Vinci Xi Surgical System, REF: IS4000",
      "product_quantity": "9",
      "reason_for_recall": "Preventative maintenance data was used to identify instrument Arms, part of surgical systems, that allow surgeons to perform surgical tasks, that may be at risk of insertion ballscrew failure, allowing for uncontrolled carriage movement along the insertion axis, which may cause tissue injury.",
      "recall_initiation_date": "20230124",
      "center_classification_date": "20230316",
      "report_date": "20230322",
      "code_info": "System REF: Serial Numbers/Cart REF/UDI-DI/Arm Serial Numbers: IS4200: SL0379, SL0069/ 380620/ 00886874115404/ 668080, 575093,  IS4000: SK1402, SK0991, SK2332, SK0744, SK0110, SK0753/380652 / 00886874110720/ 372054, 663693, 557322, 547953, 425439, 409962"
    }
  ]
}