{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "South Jordan",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87306",
      "recalling_firm": "Merit Medical Systems, Inc.",
      "address_1": "1600 W Merit Pkwy",
      "address_2": "N/A",
      "postal_code": "84095-2416",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. Consignees: VA, NE, Washington DC, MI, NY, KY, CA, TX, MD, AL, GA, UT, LA, CA, CT, PA, WA, MT and NJ.     No government consignees.    Also distributed OUS.",
      "recall_number": "Z-1262-2021",
      "product_description": "HeartSpan Transseptal Needles Catalog No. FND-019-01, Lot No. E1913644 consisting of  an outer needle cannula and an inner stylet. The needle is comprised of flexible thin walled tubing with an ergonomic hub and stopcock attached to the proximal end. The stylet consists of a solid wire that when inserted in the needle protrudes beyond the distal tip of the cannula.",
      "product_quantity": "587 Units",
      "reason_for_recall": "The labeled needle tip curvature of specific lots of transseptal needles may not match the actual needle tip curvature. In the event that the incorrect needle is unknowingly used, an unintended anatomy puncture may occur, which is likely to require medical intervention.",
      "recall_initiation_date": "20210119",
      "center_classification_date": "20210316",
      "termination_date": "20220517",
      "report_date": "20210324",
      "code_info": "Catalog Number FND-019-01, Lot Number E1913644; UDI: (01)00884450489345"
    }
  ]
}