  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82582",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK",
      "recall_number": "Z-1262-2019",
      "product_description": "NM/CT 850    Product Usage:  The GE NM/CT 850 system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography component which is intended specifically for enabling attenuation correction and anatomical localization of SPECT images.",
      "product_quantity": "31 units",
      "reason_for_recall": "During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.",
      "recall_initiation_date": "20190320",
      "center_classification_date": "20190502",
      "termination_date": "20201005",
      "report_date": "20190508",
      "code_info": "Mfg. Lot or Serial # System ID Model, Catalogue or Code #  85EX67018 4740081NM11 H3907AD  85EX67001 203789NM850 H3907AD  850X61005 319768NMCT H3907AD  850Y61006 319768NMH988 H3907AD  85EX67022 847618NM H3907AD  85EX67019 773975NMCT H3907AD  85EX67008 779696NM850 H3907AD  85EX67017 318212NM H3907AD  85EX67012 781729NM H3907AD  85EX67015 SHIPPED (AO 1623723) H3907AD  85EX67020 SHIPPED (GON 4745963) H3907AD  85EX67023 SHIPPED (GON 4704278) H3907AD  85EX67024 SHIPPED (GON 4710759) H3907AD  85EY67025 SHIPPED (AO 1638499) H3907AD  85EY67026 SHIPPED (AO 1638498) H3907AD  85EY67028 SHIPPED (AO 1638502) H3907AD  85EX67009 513732NM 5782013  85EX67002 419289D850 H3907AD  85EX67016 972969NM H3907AD  85EX67004 571472NM850 H3907AD  85EX67021 SHIPPED (GON 4640067) H3907AD  85EX67010 804289NM850A H3907AD  85EX67011 804281NM850A H3907AD  85EX67005 715717NMCT H3907AD  85EX67013 920456NM H3907AD  850X61004 030601NU06 H3907AD  850X61002 NS0001 H3907AD  NT3BG1900001BH NS0401 5376204-50-1  85EX67003 PL1692NM05 H3907AD  85EX67014 PPR59076 H3907AD  850Y61007 SK1082NM02 H3907AD"
    }
  ]
}