{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Carpinteria",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70558",
      "recalling_firm": "Dako North America Inc.",
      "address_1": "6392 Via Real",
      "address_2": "N/A",
      "postal_code": "93013-2921",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of:  AT, AU, BE, CA, CH, DE, DK, ES, FR, GB, IE, IT, JP, NL, NO, PL, and SE.",
      "recall_number": "Z-1261-2015",
      "product_description": "Autostainer Link 48 (AS480), Autostainer Plus Link Instrument (AS100), Autostainer (S3400), and Autostainer Plus (S3800). Automated slide stainer for in vitro diagnostic use.",
      "product_quantity": "409",
      "reason_for_recall": "A false negative result affecting the diagnosis may occur. A defect in the syringe assembly stopcock component may, in some circumstances, cause leakage of excess buffer onto slide location 34 on the AS480, or slide locations 35 and 36 on the AS100, S3400, and S3800.  Affected dates are from 12/2013-08/2014.",
      "recall_initiation_date": "20150212",
      "center_classification_date": "20150310",
      "termination_date": "20160211",
      "report_date": "20150318",
      "code_info": "992640 Rev. 02, 992085 Rev.no H, 992625 Rev.no F, 992326 Rev.no C    Affected dates: 12/2013-08/2014    Each device is labeled with a unique serial number.    Catalog/Model Number:  - AS480  - S3800  - S3400  - AS100"
    }
  ]
}