  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tempe",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91638",
      "recalling_firm": "Bard Peripheral Vascular Inc",
      "address_1": "1625 W 3rd St Bldg 1",
      "address_2": "N/A",
      "postal_code": "85281-2438",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) includes states of: FL, MD, OH, CO, PA, TX, NC, CA, ME, MO, NV, MI, IL, NY, VA, AZ, NH, LA, WI, KY, MA, WA, CT, SD, OK, DC, IA, AR, GA, NE, NM, IN, NJ, MN, PR, AL, AK, SC, KS, OR, ID, VT, MT, WV, ND, TN, MS, HI, WY, RI and OUS countries of: Canada, Korea, Thailand and Taiwan.  OUS: AR, AU, BE, CA, CN, CO, HK, IN, JP, KR, TH, TW",
      "recall_number": "Z-1260-2023",
      "product_description": "Bard Mission Disposable Core Biopsy Instrument Kit: 16G x 10CM, REF: 1610MSK; 18G x 16CM, REF: 1816MSK",
      "product_quantity": "38,635",
      "reason_for_recall": "The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if used despite incompatibility.",
      "recall_initiation_date": "20230228",
      "center_classification_date": "20230316",
      "report_date": "20230322",
      "code_info": "REF/UDI-DI/Lot(Expiration Date): 1610MSK/00801741142567/1445630(11/28/2024), 1449980(12/28/2024), 1451134(1/28/2025);  1816MSK/00801741097065/1447510(12/28/2024), 1447511(12/28/2024), 1447512(12/28/2024), 1447513(12/28/2024), 1448118(12/28/2024), 1448119(12/28/2024), 1448120(12/28/2024), 1448121(12/28/2024), 1448743(12/28/2024), 1450413(12/28/2024), 1450414(12/28/2024), 1451075(1/28/2025), 1451076(1/28/2025), 1451621(1/28/2025), 1451622(1/28/2025), 1451625(1/28/2025), 1451626(1/28/2025), 1453516(1/28/2025), 1453517(1/28/2025), 1454295(1/28/2025), 1456850(2/28/2025), 1456856(2/28/2025), 1457030(2/28/2025)"
    }
  ]
}