  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The  oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial  light for six years prior to inert packaging and sterilization of the device.",
      "address_2": "",
      "product_quantity": "3638 units",
      "code_info": "Item Number  Item Number  681600016  681600019  681600022  681601016  681602016  681602022  681603016  681603019  681604009  681604011  681604013  681604016  681604019  681604022  681605009  681605019",
      "center_classification_date": "20140320",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of   Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.",
      "state": "IN",
      "product_description": "Apollo\" Knee System:  PS TIB INS SZ0/16MM APOLL  PS TIB INS SZ0/19MM APOLL  PS TIB INS SZ0/22MM APOLL  PS TIB INS SZ1/16MM APOLL  PS TIB INS SZ2/16M APOLLO  PS TIB INS SZ2/22MM APOLL  PS TIB INS SZ3/16MM APOLL  PS TIB INS SZ3/19MM APOLL  PS TIB INS SZ4/9MM APOLLO  PS TIB INS SZ4/11MM APOLL  PS TIB INS SZ4/13MM APOLL  PS TIB INS SZ4/16MM APOLL  PS TIB INS SZ4/19MM APOLL  PS TIB INS SZ4/22MM APOLL  PS TIB INS SZ5/9MM APOLLO  PS TIB INS SZ5/19MM APOLL",
      "report_date": "20140326",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer, Inc.",
      "recall_number": "Z-1260-2014",
      "initial_firm_notification": "",
      "product_type": "Devices",
      "event_id": "67503",
      "termination_date": "20150324",
      "more_code_info": "",
      "recall_initiation_date": "20140225",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}