{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90113",
      "recalling_firm": "Tandem Diabetes Care Inc",
      "address_1": "11075 Roselle St",
      "address_2": "",
      "postal_code": "92121-1204",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "International distribution in the country of Australia.",
      "recall_number": "Z-1258-2022",
      "product_description": "t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technology",
      "product_quantity": "5,206",
      "reason_for_recall": "Insulin pumps may display unexpected fluctuation of the remaining battery life so ensure pump battery life does not fall below 25% remaining power. If the battery life drops to less than 5%, insulin delivery will continue for 30 minutes and then the pump will power off and insulin delivery will stop. If the battery reaches 1%, then insulin delivery will stop, which could lead to hyperglycemia.",
      "recall_initiation_date": "20200910",
      "center_classification_date": "20220610",
      "report_date": "20220622",
      "code_info": "Dexcom G5 - All Serial Numbers; Basal-IQ Technology - Software Versions 6.3.01 and Earlier",
      "more_code_info": ""
    }
  ]
}