  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82582",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK",
      "recall_number": "Z-1257-2019",
      "product_description": "Discovery NM 630    Product Usage:  The Discovery NM 630 is an all-purpose dual detector nuclear imaging system intended for general nuclear medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The main components of the Discovery NM 630 system are: NM Gantry with Dual detector heads, patient table, Remote Control Unit and NM operation console.",
      "product_quantity": "57 units",
      "reason_for_recall": "During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.",
      "recall_initiation_date": "20190320",
      "center_classification_date": "20190502",
      "termination_date": "20201005",
      "report_date": "20190508",
      "code_info": "Mfg. Lot or Serial # System ID Model, Catalogue or Code #  63GY50074 334287NM630 H3101RH  63GX50024 501450D630 H3101RH  630X22925 310657D630 H3100JY  630X22921 951682NMDISC H3100JY  63GY50067 719776DISCNM H3101RH  63GX50001 863494NM H3101RH  630X22923 904296NM H3100JY  63GX50037 305223D630A H3101RH  63GX50060 305223D630B H3101RH  63GX50034 478289DIS H3101RH  63GX50014 618998NM H3101RH  63GX50019 270745NM1 H3101RH  63GY50078 269966NMDISC H3101RH  63GX50021 906225CARDNM2 H3101RH  63GX50022 906225CARDNM1 H3101RH  63GX50018 269637NM630 H3101RH  63GX50009 NM555658 H3101RH  63GY50080 4695517NM11 H3101RH  63GY50077 919954NMDISC H3101RH  63GX50043 828286RHD630 H3101RH  63GX50020 402481D630C H3101RH  63GX50040 973450CARD630 H3101RH  63GX50054 973450D630 H3101RH  63GX50008 856641NM H3101RH  63GY50079 SHIPPED (GON 4722927) H3101RH  63GX50061 702962NM1 H3101RH  63GX50035 702962NM2 5376204-31  63GX50050 702388DISC H3101RH  63GX50062 702724DISCNM H3101RH  63GX50032 513861NM H3101RH  63GX50055 513861NM1 H3101RH  63GX50013 740MGD630 H3101RH  63GX50047 440743D630 H3101RH  63GX50002 419251NM H3101RH  63GX50025 330841DISCNM H3101RH  63GX50033 330841DISCNM2 H3101RH  63GX50038 918756D630 H3101RH  63GX50044 803865D630 H3101RH  63GY50068 854529NMDISC H3101RH  63GY50069 901516NM 5376204-31  63GX50016 615873NM H3101RH  63GY50075 SHIPPED (GON 4561460) H3101RH  63GX50012 832505D630 H3101RH  63GX50005 713440NM630B H3101RH  63GX50029 281MHNE630 H3101RH  630X22919 972258NM630 H3100JY  63GX50046 210351NM H3101RH  63GX50049 281MWL630B H3101RH  63GX50039 281MWL630A H3101RH  63GX50030 434243DIS H3101RH  63GX50011 LGMONT630 H3101RH  63GX50017 540316NM630 5791570  630X22922 804764D630C H3100JY  630X22924 UESS01NU03 H3100JY  63GX50048 418404NU02 H3101RH  63GX50041 083026202460918 H3101RH  NT2PG1800001BH NT8003 5376204-31-1"
    }
  ]
}