  "results": [
    {
      "status": "Terminated",
      "city": "East Walpole",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76285",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc",
      "address_1": "333 Coney St",
      "address_2": "N/A",
      "postal_code": "02032-1516",
      "voluntary_mandated": "N/A",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Nationwide.  Afghanistan, Albania, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium,  Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark,  Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Polynesia, Georgia,  Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq,  Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Latvia, Lithuania, Luxembourg, Macedonia,  Malaysia, Malta, Martinique, Mexico, Myanmar, Nepal, Netherlands, New Caledonia, New  Zealand, Nicaragua, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal,  Qatar, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia,  Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan,  Thailand, Tunisia, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay,  Vatican, Vietnam",
      "recall_number": "Z-1256-2017",
      "product_description": "Siemens ADVIA Centaur Systems Calibrator A Kits,   Cat No. 04800646, SMN 10285903 (2 pack)Cat No. 04800735, SMN 10285904 (6 pack)  Cat No. 04800840, SMN 10285905 (6 pack - REF)",
      "product_quantity": "194 kits",
      "reason_for_recall": "Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator RLU %CVs when using Calibrator A kit lots ending in 90 with the FT4 assay.  The performance of the FT4 assay when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur CP System is not affected.  Customers may continue to use Calibrator A kit lots ending in 90 for the ADVIA Centaur Systems FT3, T3, T4 and TUp assays.",
      "recall_initiation_date": "20170110",
      "center_classification_date": "20170225",
      "termination_date": "20190716",
      "report_date": "20170308",
      "code_info": "Lot Numbers:  08722A90  27210A90  43429A90      Exp.  Date: 2017/07/24",
      "more_code_info": ""
    }
  ]
}