  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The  oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial  light for six years prior to inert packaging and sterilization of the device.",
      "address_2": "",
      "product_quantity": "3638 units",
      "code_info": "Item Number  625001609  625001613  625001619  625001709  625001711  625001713  625001719  625002613  625002616  625002719  625002722  630809510  630811410  630811510  630813410  630813510  630816210  630816410  630816510  630819010  630819110  630819210  630819310  630819410  630819510  635000009  635001009  635001011  635001111  635001116  635002009  635002011  635002109  635002111  635002113  635003009  635003011  635003013  635003016  635003109  635003113  635003116  635004009  635004011  635004016  635004111  635004113  635004119  635005013  635005019  635005111  635005116  635006109  670501916  670501919  670503819  670503911  670503916",
      "center_classification_date": "20140320",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of   Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.",
      "state": "IN",
      "product_description": "Natural-Knee¿ II System  ULTRA CONG INS LT SZA/9MM  ULTRA CONG INS LT SZA/13M  ULTRA CONG INS LT SZA/19M  ULTRA CONG INS RT SZA/9MM  ULTRA CONG INS RT SZA/11M  ULTRA CONG INS RT SZA/13M  ULTRA CONG INS RT SZA/19M  ULTRA CONG INS LT SZB/13M  ULTRA CONG INS LT SZB/16M  ULTRA CONG INS RT SZB/19M  ULTRA CONG INS RT SZB/22M  NKII ROT PLTFM U/C INS, 9  NKII ROT PLTFM U/C INS, 1  NKII ROT PLTFM U/C INS, 1  NKII ROT PLTFM U/C INS, 1  NKII ROT PLTFM U/C INS, 1  NKII ROT PLTFM U/C INS, 1  NKII ROT PLTFM U/C INS, 1  NKII ROT PLTFM U/C INS, 1  NKII ROT PLTFM U/C INS, 1  NKII ROT PLTFM U/C INS, 1  NKII ROT PLTFM U/C INS, 1  NKII ROT PLTFM U/C INS, 1  NKII ROT PLTFM U/C INS, 1  NKII ROT PLTFM U/C INS, 1  ALL POLY CONG TIB LT SZ00  ALL POLY CONG TIB LT SZ0/  ALL POLY CONG TIB LT SZ0/  ALL POLY CONG TIB RT SZ0/  ALL POLY CONG TIB RT SZ0/  ALL POLY CONG TIB LT SZ1/  ALL POLY CONG TIB LT SZ1/  ALL POLY CONG TIB RT SZ1/  ALL POLY CONG TIB RT SZ1/  ALL POLY CONG TIB RT SZ1/  ALL POLY CONG TIB LT SZ2/  ALL POLY CONG TIB LT SZ2/  ALL POLY CONG TIB LT SZ2/  ALL POLY CONG TIB LT SZ2/  ALL POLY CONG TIB RT SZ2/  ALL POLY CONG TIB RT SZ2/  ALL POLY CONG TIB RT SZ2/  ALL POLY CONG TIB LT SZ3/  ALL POLY CONG TIB LT SZ3/  ALL POLY CONG TIB LT SZ3/  ALL POLY CONG TIB RT SZ3/  ALL POLY CONG TIB RT SZ3/  ALL POLY CONG TIB RT SZ3/  ALL POLY CONG TIB LT SZ4/  ALL POLY CONG TIB LT SZ4/  ALL POLY CONG TIB RT SZ4/  ALL POLY CONG TIB RT SZ4/  ALL POLY CONG TIB RT SZ5/  PS TIB INS RT SZ1/2 16MM  PS TIB INS RT SZ1/2 19MM  PS TIB INS LT SZ00/0 19MM  PS TIB INS RT SZ00/0 11MM  PS TIB INS RT SZ00/0 16MM",
      "report_date": "20140326",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer, Inc.",
      "recall_number": "Z-1255-2014",
      "initial_firm_notification": "",
      "product_type": "Devices",
      "event_id": "67503",
      "termination_date": "20150324",
      "more_code_info": "",
      "recall_initiation_date": "20140225",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}