  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76343",
      "recalling_firm": "Phadia US Inc",
      "address_1": "4169 Commercial Ave",
      "address_2": "N/A",
      "postal_code": "49002-9701",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: AR, CA, CO, FL, GA, IL, IN, KY, LA, MI, MN, MO, MT, NC, NH, NJ, NM, NY, OH, OR, PA, SD, TN, TX, WI, WV    Foreign: None    VA/DOD: None",
      "recall_number": "Z-1253-2017",
      "product_description": "EliA Sample Diluent, Article number, 83-1023-01",
      "product_quantity": "5957",
      "reason_for_recall": "Erroneous EliA test results related to problems with liquid level detection and diluent pipetting in Phadia 250 instruments, and deformed bottles.  Investigation showed problems with the liquid level detection in diluent bottle batch JJCA resulting in aspiration of no or too low volumes of liquid from the diluent bottle and from the dilution plate.",
      "recall_initiation_date": "20170130",
      "center_classification_date": "20170224",
      "termination_date": "20170807",
      "report_date": "20170308",
      "code_info": "EliA Sample Diluent, Article number, 83-1023-01, lots JM3Y, JS0B, JP87, JK5N, and JU9C",
      "more_code_info": ""
    }
  ]
}