  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Watertown",
      "state": "SD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70473",
      "recalling_firm": "Pivotal Health Solutions",
      "address_1": "724 Oakwood Road",
      "address_2": "N/A",
      "postal_code": "57201-4133",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to states of: AZ, AR, CA, GA, IL, IN, IA, KS, MI, MN, MO, NY, OH, OR, PA, UT and WA.",
      "recall_number": "Z-1250-2015",
      "product_description": "Apollo (cold) Laser Desktop Control Units, Model AP2-DT.    The Apollo IR Heat Lamp System is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, the temporary increase in local blood circulation and / or promoting relaxation of muscle.",
      "product_quantity": "35",
      "reason_for_recall": "Control units were equipped with an internal mounting kit that does not meet medical safety standards, and are conductive, increasing the risk of electric shock to the user and patient. These units were manufactured prior to Pivotal Health Solutions acquisition of the Apollo product line.",
      "recall_initiation_date": "20141217",
      "center_classification_date": "20150306",
      "termination_date": "20160309",
      "report_date": "20150318",
      "code_info": "DT-1102, 1105, 1106, 1109, 1110, 1112,1114, 1115, 1116, 1118,1119, 1120, 1121, 1121, 1122, 1123, 1124, 1125, 1126, 1127, 1128, 1129, 1130, 1131, 1132, 1134, 1135, 1136, 1137, 1138, 1139, 1302, 1303, 1304, 1309."
    }
  ]
}