  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70514",
      "recalling_firm": "GE Healthcare",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) including the states of  AK, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MO, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI,and  WV, and the countries of ARGENTINA, CANADA, CHILE, CZECH REPUBLIC, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, IRELAND, ITALY, JAPAN, NEW ZEALAND, PORTUGAL, RUSSIA, SAUDI ARABIA, SLOVAKIA, SPAIN, and UNITED KINGDOM.",
      "recall_number": "Z-1248-2015",
      "product_description": "GE Healthcare Proteus XR/a with wall stand model number 2260354.    Intended for use in generating radiographic images of human anatomy in all general purpose diagnostic procedures. This device is not intended for mammographic applications.",
      "product_quantity": "192 (131 US, 61 OUS).",
      "reason_for_recall": "Potential breakage of the steel cable which supports the bucky device in certain wall stands of Proteus XR/a X-ray imaging systems. A fall of a bucky while the system is in use could result in an injury to a patient or operator.",
      "recall_initiation_date": "20150130",
      "center_classification_date": "20150306",
      "termination_date": "20150915",
      "report_date": "20150318",
      "code_info": "Mfg Lot or Serial # System ID  00000014277HL8 503246PRO  00000014127HL5 202715PRO  00000013133HL4 206215RM1  00000013085HL6 206543HECED  00000014123HL4 208785RAD  00000014689HL4 208882GX1A  00000010591HL6 208893RAD2  00000014556HL5 209943SJRM2  00000013410HL6 214320RAD1  00000014413HL9 215918RAD1  00000014555HL7 215918RAD1  00000014684HL5 215918RAD2  00000014116HL8 216343R3  00000014276HL0 216343R4  00000013780HL2 217258SR3  00000024745HL2 217333LXR2  00000014020HL2 219284PRO1  00000014687HL8 225638R1  00000014119HL2 248652PROT3  00000014679HL5 262321RAD1  00000015410HL4 281392PRO  00000014424HL6 304257PROT  00000011358HL9 304354PROD  00000015166HL2 309655FNCR1  00000015168HL8 309655FNCR2  00000012515HL3 313369PROT  00000013910HL5 314768MRM8  00000014562HL3 314966JRM2A  00000011363HL9 318425LR1  00000013785HL1 319382PROTXR  00000014412HL1 330270PX  00000015409HL6 330438R3  00000014282HL8 330489XR3  00000014423HL8 330675PX  00000011060HL1 336768OSAB  00000010841HL5 336768OSCD  00000011055HL1 336768OSEF  00000010842HL3 336768OSGH  00000014686HI0 352401RAD3  00000011577HL4 360414SJRM4  00000015171HL2 360425LRP  00000013908HL9 402481PLRAD  00000015055HL7 404624METRO1  00000011195HL5 405447ORT  00000014283HL6 405744OSU  00000015047HL4 405949BCH  00000011580HL8 408636PROTEUS  00000015420HL3 414266RM8  00000013094HL8 419228LMP  00000014122HL6 419251RAD1  00000011198HL9 419946PRO2  00000013393HL4 425883ER  00000014128HL3 440835RAD1  00000014682HL9 440988RAD  00000015062HL3 502587JHC  00000012526HL0 502868RAD2  00000013390HL0 505272RM1  00000013610HL1 505835RM1  00000012524HL5 509684MC2  00000011359HL7 509684MCC1  00000014022HL8 517463RM3  00000013611HL9 517545RMONE  00000011199HL7 517684RM1  00000012510HL4 518292RAD2  00000012513HL8 518292RAD4  00000014683HL7 518483RAD  00000015050HL8 518583RAD7  00000012511HL2 518608RAD1  00000011177HL3 520742PROTEUS  00000014415HL4 520886PRO  00000013909HL7 570321TOMO  00000013600HL2 603352KC02  00000012509HL6 606474BFM  00000014021HL0 608647RAD  00000014115HL0 610280PROT1  00000013783HL6 610377PRO  00000013870HL1 614898PROT  00000014114HL3 615735PROT1  00000014685HL2 615781SHPROT  00000012865HL2 618242GS3  00000013791HL9 620343ER1  00000010882HL9 630213OAD4  00000012507HL0 630653OAD3  00000013781HL0 660679PROT  00000012690HL4 702434CAMP  00000015056HL5 702731RAD1  00000014419HL6 703698PRO1  00000015169HL6 706276NGR1  00000015053HL2 706886SR1  00000013596HL2 716275NEWER1  00000013868HL5 716477RAD  00000015063HL1 718262CLINIC  00000014559HL9 718830PROTEUS  00000013388HL4 740454PROT  00000014120HL0 740HOCPRO  00000015061HL5 760366ER  00000013088HL0 770270NR2  00000015762HL8 773967XRD3  00000013096HL3 775982PROT1  00000015165HL4 775982PROT4  00000015408HL8 801262RM3  00000011562HL6 801408RAD2  00000012514HL6 808974PRO  00000015178HL7 810421RAD  00000013090HL6 810484RM1  00000013086HL4 810484RM2  00000014411HL3 810484RM3  00000012536HL9 810758PROT  00000013914HL7 810985RAD  00000013400HL7 8149382O  00000011357HL1 815439RAD  00000013873HL5 816932SRM2  00000015172HL0 817447RAD  00000012517HL9 817605RAD  00000014417HL0 845887PRO  00000013608HL5 863419PRO  00000011058HL5 865425RAD1  00000011059HL3 865425RAD2  00000013591HL3 865544ER1  00000015174HL6 904819RAD  00000013913HL9 906485RAD  00000013391HL8 907264PROTEUS  00000013087HL2 908277RAD1  00000014117HL6 913791ORM4  00000012684HL7 915573RAD1  00000012686HL4 915573RAD2  00000013874HL3 919553WRAD1  00000014275HL2 925370LMRAD1  00000012521HL1 941793RAD  00000013594HL7 970521RM1  00000015421HL1 973971PED1  00000015052HL4 14308PROT1  00000015059HL9 14308PROT2  00000013387HL6 204325RAD2  00000011570HL9 306695RAD  00000011572HL5 418668PROTEUS  00000012683HL9 506466RM1A  00000014422HL0 519322PRO  00000011568HL3 519364GEN  00000013911HL3 705789RAD2  00000011579HL0 709897PROTEUS  00000014561HL5 807727RAD1  00000011354HL8 819563DPROTEUS  00000011196HL3 905468RAD  00000014416HL2 905732RAD  00000012528HL6 905852RAD  00000015058HL1 26329XR1  00000015417HL9 CS1147RX01  00000011181HL5 M4496931  00000013396HL7 CC3081XR02  00000011056HL9 HC1413XR01  00000012508HL8 HC7935XR01  00000011365HL4 HDAR01XR02  00000011362HL1 PX1401XR01  00000013409HL8 H2064RAD01  00000013790HL1 0828070086  00000014418HL8 HU1113RX01  00000013395HL9 IE1019RX02  00000011192HL2 IE1039RX01  00000012516HL1 A5109116  00000012505HL4 A5109117  00000015403HL9 A5113124  00000014017HL8 XU0005  00000013875HL0 XU0007  00000015176HL1 XU0008  00000013598HL8 0920079043  00000013593HL9 0920079044  00000014113HL5 PPR58501  00000015170HL4 PPR59009  00000015064HL9 PPR59010  00000013134HL2 RU1754RX01  00000012685HL4 SA1220RX09  00000015405HL4 SK1029RX01  00000013095HL5 9P252500  00000011355HL5 CP714307  00000014121HL8 DE658838  00000015407HL0 DEH35102  00000015057HL3 DEN01309  00000015413HL8 DEN01310  00000014552HL4 FE394804  00000010653HL5 FPG72301  00000015173HL8 MEX16906  00000012527HL8 YE151109  00000013597HL0 YE637005  00000015048HL2 00115RAD02  00000015167HL0 00115RAD03  00000011356HL3 00126RAD02  00000014690HL2 00160RAD06  00000010329HL1 00433RAD02  00000015175HL3 00832RAD02  00000012531HL0 01006RAD01  00000013392HL6 10259RAD02"
    }
  ]
}