  "results": [
    {
      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90184",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1246-2022",
      "product_description": "Artis icono floor-angiography systems developed for single and biplane diagnostic imaging and interventional procedures   SMN: 11327700",
      "product_quantity": "13 units",
      "reason_for_recall": "if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock  if a person is in contact with metallic parts of the C\u0010arm",
      "recall_initiation_date": "20220411",
      "center_classification_date": "20220607",
      "report_date": "20220615",
      "code_info": "UDI-DI: 04056869149325 Serial Number: 170026 170305 170307 170309 170310 170023 170304 170025 170051 170312 170027 170303 170053",
      "more_code_info": ""
    }
  ]
}