  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91671",
      "recalling_firm": "Maquet Cardiovascular, LLC",
      "address_1": "45 Barbour Pond Dr",
      "address_2": "N/A",
      "postal_code": "07470-2094",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of FL, IN, LA, MI, MO, NC, NV, NY, PA, TN, TX, VA, WA.",
      "recall_number": "Z-1245-2023",
      "product_description": "Fusion Bioline Vascular Graft, Part number M00201503046B0",
      "product_quantity": "1 (US)",
      "reason_for_recall": "One (1) Fusion Bioline Vascular graft from lot 25162546, which failed guideline straightness testing, was intermingled with lot 25162549 and distributed to customers. Use of the affected graft may result in reduction of blood flow or unwanted tension causing bleeding and the formation of a pseudo-aneurysm.",
      "recall_initiation_date": "20230130",
      "center_classification_date": "20230310",
      "report_date": "20230322",
      "code_info": "UDI-DI 00384409005942 Lot 25162546"
    }
  ]
}