  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90184",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1245-2022",
      "product_description": "Artis icono biplane -angiography systems developed for single and biplane diagnostic imaging and interventional procedures   SMN: 11327600",
      "product_quantity": "51 units",
      "reason_for_recall": "if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock  if a person is in contact with metallic parts of the C\u0010arm",
      "recall_initiation_date": "20220411",
      "center_classification_date": "20220607",
      "report_date": "20220615",
      "code_info": "UDI-DI: 04056869063317  Serial Number: 180062 180054 180019 180065 180056 180057 180041 180317 180034 180072 180060 180070 180048 180083 180315 180080 180301 180305 180067 180330 180040 180333 180029 180336 180316 180084 180334 180018 180046 180047 180335 180085 180035 180081 180331 180021 180321 180309 180323 180329 180043 180049 180077 180086 180053 180055 180015 180014 180066 180082 180037",
      "more_code_info": ""
    }
  ]
}