      "status": "Terminated",
      "city": "HIROSHIMA",
      "state": "N/A",
      "country": "Japan",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72991",
      "recalling_firm": "MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK",
      "address_1": "6-22, 4-CHOME, KAN-ON-SHIN-MACHI",
      "address_2": "NISHI-KU",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed in the states of Florida, New York, Ohio & Texas, and the countries of France, Germany, Japan, Italy, Korea,  & Belgium.",
      "recall_number": "Z-1244-2016",
      "product_description": "MHI-TM2000 Linear Accelerator System (Software Version 3.0.0 and after)  Intended for radiation therapy of lesions, tumors. conditions anywhere in the body where radiation therapy is indicated.",
      "product_quantity": "25 devices (4 domestically in U.S.A. and 21 internationally)",
      "reason_for_recall": "Due to Operator Console software anomaly, a change in treatment completion status of the last patient of the day may be altered from \"Completed (or Discontinued)\" to \"Untreated\" under specific conditions.",
      "recall_initiation_date": "20151215",
      "center_classification_date": "20160324",
      "termination_date": "20170420",
      "report_date": "20160330",
      "code_info": "Serial numbers: 201901, 201902, 201903, 201904, 201905, 202901, 202902, 202903, 202904, 202905, 202906, 203901, 203902, 203903, 203904, 203905, 203906, 203918, 203919, 203921, 203922, 203924, 203925, 203927 and 203929."
    }
  ]
}