      "status": "Terminated",
      "city": "East Walpole",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73052",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc",
      "address_1": "333 Coney St",
      "address_2": "N/A",
      "postal_code": "02032-1516",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution \u0013 US Nationwide , AR, AU, AU, BD, BO, BR, CA, CH, CL, CN, CO, DO, EC, EG, HK, ID, IE, IL, IN, JP, KR, LK, LY, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VE, VN, and ZA.",
      "recall_number": "Z-1243-2016",
      "product_description": "IMMULITE 2000 EstradiolTest Code E2  Catalog Number: L2KE22, L2KE22 (D), L2KE26, L2KE26 (D)  SMN: 10381178, 10702833, 10381177, 10702834    Product Usage:  For in vitro diagnostic use the IMMULITE¿2000 systems Analyzers- for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various sexual disorders",
      "product_quantity": "9334 units",
      "reason_for_recall": "The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results.",
      "recall_initiation_date": "20160113",
      "center_classification_date": "20160324",
      "termination_date": "20180108",
      "report_date": "20160330",
      "code_info": "All lots"
    }
  ]
}