  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Vernon Hills",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70220",
      "recalling_firm": "Richard Wolf Medical Instruments Corp.",
      "address_1": "353 Corporate Woods Pkwy",
      "address_2": "N/A",
      "postal_code": "60061-3110",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-1240-2015",
      "product_description": "The I-Cart is used to store, transport, and power video equipment used in medical procedures.",
      "product_quantity": "83 carts",
      "reason_for_recall": "The transformers installed on carts used for Richard Wolf medical video systems may overheat causing smoke and black plastic material to leak from the cart chassis.",
      "recall_initiation_date": "20141210",
      "center_classification_date": "20150303",
      "termination_date": "20170731",
      "report_date": "20150311",
      "code_info": "(NOTE: The listed serial numbers are for the carts manufactured and include those that were distributed.)     1) Part Number: 31113.801; Serial Numbers: 1306000, 1306007, 1306014, 1306021, 1306103, 1306001, 1306008, 1306015, 1306022, 1306104, 1306002, 1306009, 1306016, 1306023, 1306105, 1306003, 1306010, 1306017, 1306024, 1307100, 1306004, 1306011, 1306018, 1306100,   1307101, 1306005, 1306012, 1306019, 1306101, 1307102, 1306006, 1306013, 1306020, 1306102, 1317103.     2) Part Number: 31113.701; Serial Numbers: 11001, 11011, 11020, 11030, 11039, 11048, 11058, 11002, 11012, 11021, 11031, 11040, 11049, 11059, 11003, 11013, 11023, 11032, 11041, 11050, 11060, 11005, 11014, 11024, 11033, 11042, 11051, 11061, 11006, 11015, 11025, 11034, 11043, 11052, 11063, 11007, 11016, 11026, 11035, 11044, 11053, 11064, 11008, 11017, 11027, 11036, 11045, 11055, 11065, 11009, 11018, 11028, 11037, 11046, 11056, 11066, 11010, 11019, 11029, 11038, 11047, 11057, 11067."
    }
  ]
}