      "country": "United States",
      "city": "North Haven",
      "address_1": "60 Middletown Ave",
      "reason_for_recall": "Product reportedly shutting down during use.",
      "address_2": "",
      "product_quantity": "2121",
      "code_info": "SIGPHANDLE-UDI #10884521543782  Serial#  C17AAA0001  C17AAA0002  C17AAA0003  C17AAA0004  C17AAA0005  C17AAA0006  C17AAA0007  C17AAA0008  C17AAA0009  C17AAA0010  C17AAA0011  C17AAA0012  C17AAA0013  C17AAA0014  C17AAA0015  C17AAA0017  C17AAA0018  C17AAA0019  C17AAA0020  C17AAA0021  C17AAA0022  C17AAA0023  C17AAA0024  C17AAA0026  C17AAA0027  C17AAA0028  C17AAA0029  C17AAA0030  C17AAA0031  C17AAA0032  C17AAA0033  C17AAA0034  C17AAA0035  C17AAA0036  C17AAA0037  C17AAA0038  C17AAA0039  C17AAA0040  C17AAA0042  C17AAA0043  C17AAA0044  C17AAA0045  C17AAA0046  C17AAA0047  C17AAA0048  C17AAA0049  C17AAA0050  C17AAA0051  C17AAA0052  C17AAA0053  C17AAA0054  C17AAA0055  C17AAA0056  C17AAA0057  C17AAA0058  C17AAA0059  C17AAB0002  C17AAB0003  C17AAB0004  C17AAB0005  C17AAB0006  C17AAB0007  C17AAB0008  C17AAB0009  C17AAB0010  C17AAB0011  C17AAB0012  C17AAB0013  C17AAB0014  C17AAB0015  C17AAB0016  C17AAB0017  C17AAB0018  C17AAB0019  C17AAB0020  C17AAB0021  C17AAB0022  C17AAB0023  C17AAB0024  C17AAB0025  C17AAB0026  C17AAB0027  C17AAB0028  C17AAB0029  C17AAB0030  C17AAB0031  C17AAB0032  C17AAB0033  C17AAB0034  C17AAB0035  C17AAB0036  C17AAB0037  C17AAB0038  C17AAB0039  C17AAB0040  C17AAB0041  C17AAB0042  C17AAB0043  C17AAB0044  C17AAB0045  C17AAB0046  C17AAB0047  C17AAB0048  C17AAB0049  C17AAB0050  C17AAB0051  C17AAB0052  C17AAB0053  C17AAB0054  C17AAB0055  C17AAB0056  C17AAB0057  C17AAB0058  C17AAB0059  C17AAB0060  C17AAB0061  C17AAB0062  C17AAB0063  C17AAB0064  C17AAB0065  C17AAB0066  C17AAB0069  C17AAB0070  C17AAB0072  C17AAB0073  C17AAB0074  C17AAB0075  C17AAB0076  C17AAB0077  C17AAB0078  C17AAB0079  C17AAB0080  C17AAB0081  C17AAB0082  C17AAB0083  C17AAB0084  C17AAB0085  C17AAB0086  C17AAB0087  C17AAB0088  C17AAB0089  C17AAB0090  C17AAB0091  C17AAB0092  C17AAB0093  C17AAB0094  C17AAB0095  C17AAB0096  C17AAB0097  C17AAB0098  C17AAB0099  C17AAB0101  C17AAB0102  C17AAB0103  C17AAB0104  C17AAB0105  C17AAB0107  C17AAB0108  C17AAB0109  C17AAB0110  C17AAB0111  C17AAB0112  C17AAB0113  C17AAB0114  C17AAB0115  C17AAB0116  C17AAB0118  C17AAB0120  C17AAB0121  C17AAB0122  C17AAB0123  C17AAB0124  C17AAB0125  C17AAB0126  C17AAB0127  C17AAB0128  C17AAB0129  C17AAB0130  C17AAB0131  C17AAB0132  C17AAB0133  C17AAB0134  C17AAB0135  C17AAB0136  C17AAB0137  C17AAB0138  C17AAB0139  C17AAB0140  C17AAB0141  C17AAB0142  C17AAB0143  C17AAB0145  C17AAB0146  C17AAB0147  C17AAB0148  C17AAB0149  C17AAB0150  C17AAB0151  C17AAB0152  C17AAB0153  C17AAB0155  C17AAB0156  C17AAB0157  C17AAB0158  C17AAB0159  C17AAB0160  C17AAB0161  C17AAB0162  C17AAB0163  C17AAB0164  C17AAB0165  C17AAB0166  C17AAB0167  C17AAB0168  C17AAB0169  C17AAB0170  C17AAB0171  C17AAB0172  C17AAB0173  C17AAB0174  C17AAB0175  C17AAB0176  C17AAB0177  C17AAB0178  C17AAB0179  C17AAB0180  C17AAB0181  C17AAB0182  C17AAB0183  C17AAB0184  C17AAB0185  C17AAB0186  C17AAB0187  C17AAB0188  C17AAB0189  C17AAB0190  C17AAB0191  C17AAB0192  C17AAB0193  C17AAB0194  C17AAB0195  C17AAB0196  C17AAB0197  C17AAB0198  C17AAB0199  C17AAB0200  C17AAB0201  C17AAB0202  C17AAB0203  C17AAB0204  C17AAB0205  C17AAB0206  C17AAB0207  C17AAB0208  C17AAB0209  C17AAB0210  C17AAB0211  C17AAB0212  C17AAB0213  C17AAB0214  C17AAB0215  C17AAB0216  C17AAB0217  C17AAB0218  C17AAB0219  C17AAB0220  C17AAB0221  C17AAB0222  C17AAB0223  C17AAB0224  C17AAB0225  C17AAB0226  C17AAB0227  C17AAB0228  C17AAB0230  C17AAB0231  C17AAB0232  C17AAC0001  C17AAC0002  C17AAC0003  C17AAC0004  C17AAC0005  C17AAC0006  C17AAC0007  C17AAC0008  C17AAC0009  C17AAC0010  C17AAC0011  C17AAC0012  C17AAC0013  C17AAC0014  C17AAC0015  C17AAC0016  C17AAC0017  C17AAC0018  C17AAC0019  C17AAC0020  C17AAC0022  C17AAC0023  C17AAC0024  C17AAC0025  C17AAC0026  C17AAC0027  C17AAC0028  C17AAC0029  C17AAC0030  C17AAC0032  C17AAC0033  C17AAC0034  C17AAC0035  C17AAC0036  C17AAC0037  C17AAC0039  C17AAC0040  C17AAC0041  C17AAC0042  C17AAC0043  C17AAC0044  C17AAC0045  C17AAC0046  C17AAC0047  C17AAC0048  C17AAC0049  C17AAC0050  C17AAC0051  C17AAC0052  C17AAC0053  C17AAC0054  C17AAC0055  C17AAC0056  C17AAC0057  C17AAC0058  C17AAC0059  C17AAC0060  C17AAC0061  C17AAC0062  C17AAC0063  C17AAC0064  C17AAC0065  C17AAC0066  C17AAC0067  C17AAC0068  C17AAC0069  C17AAC0070  C17AAC0071  C17AAC0072  C17AAC0073  C17AAC0074  C17AAC0075  C17AAC0076  C17AAC0077  C17AAC0078  C17AAC0079  C17AAC0080  C17AAC0081  C17AAC0082  C17AAC0083  C17AAC0084  C17AAC0085  C17AAC0086  C17AAC0087  C17AAC0088  C17AAC0089  C17AAC0090  C17AAC0091  C17AAC0092  C17AAC0093  C17AAC0094  C17AAC0095  C17AAC0096  C17AAC0097  C17AAC0098  C17AAC0099  C17AAC0100  C17AAC0101  C17AAC0102  C17AAC0103  C17AAC0104  C17AAC0105  C17AAC0106  C17AAC0107  C17AAC0108  C17AAC0109  C17AAC0110  C17AAC0111  C17AAC0112  C17AAC0113  C17AAC0115  C17AAC0116  C17AAC0118  C17AAC0119  C17AAC0120  C17AAC0121  C17AAC0122  C17AAC0123  C17AAC0124  C17AAC0125  C17AAC0126  C17AAC0127  C17AAC0128  C17AAC0129  C17AAC0130  C17AAC0131  C17AAC0132  C17AAC0133  C17AAC0134  C17AAC0135  C17AAC0136  C17AAC0137  C17AAC0139  C17AAC0140  C17AAC0141  C17AAC0142  C17AAC0143  C17AAC0144  C17AAC0145  C17AAC0146  C17AAC0147  C17AAC0148  C17AAC0149  C17AAC0150  C17AAC0151  C17AAC0152  C17AAC0153  C17AAC0154  C17AAC0155  C17AAC0156  C17AAC0157  C17AAC0158  C17AAC0159  C17AAC0160  C17AAC0161  C17AAC0162  C17AAC0163  C17AAC0164  C17AAC0165  C17AAC0166  C17AAC0167  C17AAC0168  C17AAC0169  C17AAC0170  C17AAC0171  C17AAC0172  C17AAC0173  C17AAC0174  C17AAC0175  C17AAC0176  C17AAC0177  C17AAC0178  C17AAC0179  C17AAC0180  C17AAC0181  C17AAC0182  C17AAC0183  C17AAC0184  C17AAC0185  C17AAC0186  C17AAC0187  C17AAC0188  C17AAC0189  C17AAC0190  C17AAC0191  C17AAC0192  C17AAC0193  C17AAC0194  C17AAC0195  C17AAC0196  C17AAC0197  C17AAC0198  C17AAC0199  C17AAC0200  C17AAC0201  C17AAC0202  C17AAC0203  C17AAC0204  C17AAC0205  C17AAC0206  C17AAC0207  C17AAC0208  C17AAC0209  C17AAC0210  C17AAC0211  C17AAC0212  C17AAC0213  C17AAC0214  C17AAC0215  C17AAC0216  C17AAC0217  C17AAC0218  C17AAC0219  C17AAC0220  C17AAC0221  C17AAC0222  C17AAC0223  C17AAC0224  C17AAC0225  C17AAC0226  C17AAC0227  C17AAC0230  C17AAC0231  C17AAC0232  C17AAC0233  C17AAC0234  C17AAC0235  C17AAC0236  C17AAC0237  C17AAC0238  C17AAC0239  C17AAC0240  C17AAC0241  C17AAC0242  C17AAC0243  C17AAC0244  C17AAC0245  C17AAC0246  C17AAC0247  C17AAC0248  C17AAC0249  C17AAC0250  C17AAC0251  C17AAC0252  C17AAC0253  C17AAC0254  C17AAC0255  C17AAC0256  C17AAC0257  C17AAC0258  C17AAC0259  C17AAC0260  C17AAC0261  C17AAC0262  C17AAC0263  C17AAC0264  C17AAC0265  C17AAC0266  C17AAC0267  C17AAC0268  C17AAC0269  C17AAC0270  C17AAC0271  C17AAC0272  C17AAC0273  C17AAC0274  C17AAC0275  C17AAC0276  C17AAC0277  C17AAC0280  C17AAC0281  C17AAC0282  C17AAC0283  C17AAC0284  C17AAC0285  C17AAC0286  C17AAC0287  C17AAC0288  C17AAC0289  C17AAC0290  C17AAC0291  C17AAC0292  C17AAC0293  C17AAC0294  C17AAC0295  C17AAC0296  C17AAC0297  C17AAC0298  C17AAC0299  C17AAC0300  C17AAC0301  C17AAC0302  C17AAC0303  C17AAC0304  C17AAC0305  C17AAC0306  C17AAC0307  C17AAC0308  C17AAC0309  C17AAC0310  C17AAC0311  C17AAC0312  C17AAC0313  C17AAC0314  C17AAC0315  C17AAC0316  C17AAC0317  C17AAC0318  C17AAC0319  C17AAC0320  C17AAD0002  C17AAD0003  C17AAD0004  C17AAD0005  C17AAD0006  C17AAD0008  C17AAD0009  C17AAD0010  C17AAD0011  C17AAD0012  C17AAD0013  C17AAD0014  C17AAD0015  C17AAD0016  C17AAD0017  C17AAD0018  C17AAD0019  C17AAD0020  C17AAD0021  C17AAD0022  C17AAD0023  C17AAD0024  C17AAD0025  C17AAD0026  C17AAD0027  C17AAD0028  C17AAD0030  C17AAD0031  C17AAD0032  C17AAD0033  C17AAD0034  C17AAD0035  C17AAD0037  C17AAD0038  C17AAD0039  C17AAD0040  C17AAD0041  C17AAD0043  C17AAD0044  C17AAD0046  C17AAD0048  C17AAD0049  C17AAD0050  C17AAD0051  C17AAD0052  C17AAD0053  C17AAD0054  C17AAD0055  C17AAD0056  C17AAD0057  C17AAD0058  C17AAD0059  C17AAD0060  C17AAD0061  C17AAD0062  C17AAD0063  C17AAD0064  C17AAD0065  C17AAD0066  C17AAD0067  C17AAD0068  C17AAD0069  C17AAD0070  C17AAD0071  C17AAD0072  C17AAD0073  C17AAD0074  C17AAD0075  C17AAD0076  C17AAD0077  C17AAD0078  C17AAD0079  C17AAD0081  C17AAD0082  C17AAD0083  C17AAD0084  C17AAD0085  C17AAD0086  C17AAD0087  C17AAD0088  C17AAD0089  C17AAD0090  C17AAD0091  C17AAD0092  C17AAD0093  C17AAD0094  C17AAD0095  C17AAD0096  C17AAD0097  C17AAD0098  C17AAD0099  C17AAD0100  C17AAD0101  C17AAD0102  C17AAD0103  C17AAD0104  C17AAD0105  C17AAD0106  C17AAD0108  C17AAD0110  C17AAD0111  C17AAD0112  C17AAD0113  C17AAD0114  C17AAD0115  C17AAD0116  C17AAD0117  C17AAD0118  C17AAD0119  C17AAD0120  C17AAD0121  C17AAD0122  C17AAD0123  C17AAD0124  C17AAD0125  C17AAD0126  C17AAD0127  C17AAD0128  C17AAD0129  C17AAD0130  C17AAD0131  C17AAD0133  C17AAD0134  C17AAD0135  C17AAD0136  C17AAD0137  C17AAD0138  C17AAD0139  C17AAD0140  C17AAD0141  C17AAD0142  C17AAD0143  C17AAD0144  C17AAD0145  C17AAD0146  C17AAD0148  C17AAD0149  C17AAD0151  C17AAD0152  C17AAD0153  C17AAD0154  C17AAD0155  C17AAD0156  C17AAD0157  C17AAD0158  C17AAD0159  C17AAD0160  C17AAD0161  C17AAD0162  C17AAD0163  C17AAD0164  C17AAD0165  C17AAD0166  C17AAD0167  C17AAD0168  C17AAD0169  C17AAD0170  C17AAD0172  C17AAD0173  C17AAD0174  C17AAD0175  C17AAD0176  C17AAD0177  C17AAD0178  C17AAD0179  C17AAD0180  C17AAD0181  C17AAD0182  C17AAD0184  C17AAD0185  C17AAD0186  C17AAD0187  C17AAD0188  C17AAD0189  C17AAD0190  C17AAD0191  C17AAD0192  C17AAD0194  C17AAD0195  C17AAD0196  C17AAD0197  C17AAD0198  C17AAD0200  C17AAD0201  C17AAD0202  C17AAD0203  C17AAD0204  C17AAD0205  C17AAD0207  C17AAD0208  C17AAD0209  C17AAD0210  C17AAD0211  C17AAD0212  C17AAD0213  C17AAD0214  C17AAD0215  C17AAD0216  C17AAD0217  C17AAD0218  C17AAD0219  C17AAD0220  C17AAD0221  C17AAD0222  C17AAD0223  C17AAD0224  C17AAD0225  C17AAD0227  C17AAD0228  C17AAD0229  C17AAD0230  C17AAD0231  C17AAD0232  C17AAD0233  C17AAD0234  C17AAD0235  C17AAD0236  C17AAD0237  C17AAD0238  C17AAD0239  C17AAD0240  C17AAD0241  C17AAD0242  C17AAD0243  C17AAD0244  C17AAD0245  C17AAD0246  C17AAD0249  C17AAD0256  C17AAD0257  C17AAD0258  C17AAD0259  C17AAD0261  C17AAD0263  C17AAD0264  C17AAD0265  C17AAD0266  C17AAD0273  C17AAE0001  C17AAE0002  C17AAE0003  C17AAE0004  C17AAE0005  C17AAE0006  C17AAE0007  C17AAE0008  C17AAE0009  C17AAE0010  C17AAE0011  C17AAE0012  C17AAE0013  C17AAE0014  C17AAE0015  C17AAE0017  C17AAE0018  C17AAE0019  C17AAE0020  C17AAE0021  C17AAE0022  C17AAE0023  C17AAE0024  C17AAE0025  C17AAE0026  C17AAE0027  C17AAE0028  C17AAE0029  C17AAE0030  C17AAE0031  C17AAE0032  C17AAE0033  C17AAE0034  C17AAE0035  C17AAE0036  C17AAE0037  C17AAE0038  C17AAE0039  C17AAE0040  C17AAE0041  C17AAE0042  C17AAE0043  C17AAE0044  C17AAE0045  C17AAE0046  C17AAE0047  C17AAE0048  C17AAE0049  C17AAE0051  C17AAE0052  C17AAE0053  C17AAE0054  C17AAE0055  C17AAE0056  C17AAE0057  C17AAE0058  C17AAE0060  C17AAE0061  C17AAE0062  C17AAE0063  C17AAE0064  C17AAE0065  C17AAE0066  C17AAE0067  C17AAE0068  C17AAE0069  C17AAE0070  C17AAE0072  C17AAE0073  C17AAE0074  C17AAE0075  C17AAE0076  C17AAE0078  C17AAE0079  C17AAE0080  C17AAE0081  C17AAE0082  C17AAE0083  C17AAE0084  C17AAE0085  C17AAE0087  C17AAE0088  C17AAE0089  C17AAE0090  C17AAE0091  C17AAE0092  C17AAE0093  C17AAE0094  C17AAE0095  C17AAE0096  C17AAE0098  C17AAE0099  C17AAE0100  C17AAE0101  C17AAE0102  C17AAE0103  C17AAE0104  C17AAE0105  C17AAE0106  C17AAE0107  C17AAE0108  C17AAE0109  C17AAE0112  C17AAE0113  C17AAE0114  C17AAE0116  C17AAE0117  C17AAE0118  C17AAE0119  C17AAE0120  C17AAE0121  C17AAE0122  C17AAE0123  C17AAE0124  C17AAE0125  C17AAE0126  C17AAE0127  C17AAE0128  C17AAE0130  C17AAE0131  C17AAE0132  C17AAE0133  C17AAE0134  C17AAE0135  C17AAE0136  C17AAE0137  C17AAE0138  C17AAE0139  C17AAE0140  C17AAE0142  C17AAE0143  C17AAE0144  C17AAE0145  C17AAE0146  C17AAE0147  C17AAE0148  C17AAE0149  C17AAE0150  C17AAE0152  C17AAE0153  C17AAE0154  C17AAE0155  C17AAE0156  C17AAE0157  C17AAE0158  C17AAE0159  C17AAE0160  C17AAE0161  C17AAE0162  C17AAE0163  C17AAE0164  C17AAE0165  C17AAE0166  C17AAE0167  C17AAE0168  C17AAE0169  C17AAE0171  C17AAE0173  C17AAE0174  C17AAE0175  C17AAE0176  C17AAE0177  C17AAE0178  C17AAE0179  C17AAE0180  C17AAE0181  C17AAE0182  C17AAE0183  C17AAE0184  C17AAE0185  C17AAE0186  C17AAE0187  C17AAE0188  C17AAE0189  C17AAE0190  C17AAE0191  C17AAE0192  C17AAE0193  C17AAE0195  C17AAE0196  C17AAE0197  C17AAE0198  C17AAE0199  C17AAE0200  C17AAE0201  C17AAE0202  C17AAE0203  C17AAE0204  C17AAE0206  C17AAE0207  C17AAE0208  C17AAE0209  C17AAE0210  C17AAE0211  C17AAE0212  C17AAE0213  C17AAE0214  C17AAE0215  C17AAE0216  C17AAE0217  C17AAE0218  C17AAE0219  C17AAE0220  C17AAE0221  C17AAE0222  C17AAE0223  C17AAE0224  C17AAE0225  C17AAE0226  C17AAE0227  C17AAE0228  C17AAE0229  C17AAE0230  C17AAE0231  C17AAE0232  C17AAE0233  C17AAE0234  C17AAE0235  C17AAE0236  C17AAE0237  C17AAE0238  C17AAE0240  C17AAF0029  C17AAF0033  C17AAF0034  C17AAF0035  C17AAF0036  C17AAF0037  C17AAF0038  C17AAF0039  C17AAF0040  C17AAF0066  C17AAF0067  C17AAF0071  C17AAF0072  C17AAF0074  C17AAF0075  C17AAF0076  C17AAF0077  C17AAF0078  C17AAF0079  C17AAF0080  C17AAF0081  C17AAF0082  C17AAF0083  C17AAF0084  C17AAF0085  C17AAF0086  C17AAF0087  C17AAF0088  C17AAF0089  C17AAF0090  C17AAF0091  C17AAF0092  C17AAF0093  C17AAF0094  C17AAF0095  C17AAF0097  C17AAF0098  C17AAF0099  C17AAF0100  C17AAF0101  C17AAF0102  C17AAF0103  C17AAF0104  C17AAF0107  C17AAF0108  C17AAF0109  C17AAF0110  C17AAF0111  C17AAF0113  C17AAF0114  C17AAF0115  C17AAF0116  C17AAF0117  C17AAF0118  C17AAF0119  C17AAF0120  C17AAF0121  C17AAF0122  C17AAF0123  C17AAF0124  C17AAF0125  C17AAF0126  C17AAF0127  C17AAF0128  C17AAF0129  C17AAF0130  C17AAF0131  C17AAF0132  C17AAF0133  C17AAF0134  C17AAF0135  C17AAF0136  C17AAF0137  C17AAF0138  C17AAF0139  C17AAF0140  C17AAF0141  C17AAF0142  C17AAF0143  C17AAF0144  C17AAF0145  C17AAF0146  C17AAF0147  C17AAF0149  C17AAF0150  C17AAF0151  C17AAF0152  C17AAF0153  C17AAF0154  C17AAF0155  C17AAF0156  C17AAF0157  C17AAF0158  C17AAF0159  C17AAF0160  C17AAF0161  C17AAF0162  C17AAF0163  C17AAF0164  C17AAF0165  C17AAF0166  C17AAF0167  C17AAF0168  C17AAF0169  C17AAF0170  C17AAF0171  C17AAF0172  C17AAF0173  C17AAF0174  C17AAF0175  C17AAF0176  C17AAF0177  C17AAF0178  C17AAF0179  C17AAF0180  C17AAF0181  C17AAF0182  C17AAF0183  C17AAF0184  C17AAF0185  C17AAF0186  C17AAF0187  C17AAF0188  C17AAF0189  C17AAF0190  C17AAF0192  C17AAF0193  C17AAF0194  C17AAF0195  C17AAF0196  C17AAF0197  C17AAF0198  C17AAF0199  C17AAF0200  C17AAF0201  C17AAF0202  C17AAF0203  C17AAF0204  C17AAF0205  C17AAF0206  C17AAF0207  C17AAF0208  C17AAF0209  C17AAF0210  C17AAF0211  C17AAF0212  C17AAF0213  C17AAF0215  C17AAF0216  C17AAF0217  C17AAF0218  C17AAF0219  C17AAF0220  C17AAG0001  C17AAG0002  C17AAG0003  C17AAG0004  C17AAG0005  C17AAG0006  C17AAG0007  C17AAG0008  C17AAG0009  C17AAG0010  C17AAG0011  C17AAG0012  C17AAG0013  C17AAG0014  C17AAG0016  C17AAG0017  C17AAG0018  C17AAG0019  C17AAG0021  C17AAG0022  C17AAG0023  C17AAG0024  C17AAG0025  C17AAG0026  C17AAG0027  C17AAG0028  C17AAG0029  C17AAG0030  C17AAG0031  C17AAG0032  C17AAG0033  C17AAG0035  C17AAG0036  C17AAG0037  C17AAG0038  C17AAG0039  C17AAG0040  C17AAG0041  C17AAG0042  C17AAG0043  C17AAG0044  C17AAG0045  C17AAG0047  C17AAG0048  C17AAG0049  C17AAG0050  C17AAG0051  C17AAG0052  C17AAG0053  C17AAG0054  C17AAG0055  C17AAG0056  C17AAG0057  C17AAG0059  C17AAG0060  C17AAG0061  C17AAG0062  C17AAG0063  C17AAG0064  C17AAG0065  C17AAG0066  C17AAG0069  C17AAG0070  C17AAG0071  C17AAG0072  C17AAG0073  C17AAG0074  C17AAG0075  C17AAG0077  C17AAG0078  C17AAG0079  C17AAG0080  C17AAG0081  C17AAG0082  C17AAG0083  C17AAG0084  C17AAG0085  C17AAG0086  C17AAG0087  C17AAG0088  C17AAG0089  C17AAG0090  C17AAG0091  C17AAG0092  C17AAG0093  C17AAG0094  C17AAG0095  C17AAG0096  C17AAG0098  C17AAG0099  C17AAG0100  C17AAG0101  C17AAG0102  C17AAG0103  C17AAG0104  C17AAG0105  C17AAG0106  C17AAG0107  C17AAG0108  C17AAG0109  C17AAG0110  C17AAG0111  C17AAG0112  C17AAG0113  C17AAG0115  C17AAG0116  C17AAG0117  C17AAG0118  C17AAG0119  C17AAG0120  C17AAG0121  C17AAG0122  C17AAG0123  C17AAG0125  C17AAG0126  C17AAG0127  C17AAG0128  C17AAG0129  C17AAG0130  C17AAG0131  C17AAG0132  C17AAG0133  C17AAG0134  C17AAG0135  C17AAG0136  C17AAG0137  C17AAG0138  C17AAG0139  C17AAG0140  C17AAG0141  C17AAG0142  C17AAG0143  C17AAG0144  C17AAG0145  C17AAG0146  C17AAG0147  C17AAG0148  C17AAG0149  C17AAG0151  C17AAG0152  C17AAG0153  C17AAG0154  C17AAG0155  C17AAG0156  C17AAG0157  C17AAG0158  C17AAG0159  C17AAG0160  C17AAG0161  C17AAG0162  C17AAG0163  C17AAG0164  C17AAG0165  C17AAG0166  C17AAG0167  C17AAG0168  C17AAG0169  C17AAG0170  C17AAG0171  C17AAG0172  C17AAG0173  C17AAG0174  C17AAG0175  C17AAG0176  C17AAG0177  C17AAG0178  C17AAG0179  C17AAG0180  C17AAG0181  C17AAG0182  C17AAG0183  C17AAG0184  C17AAG0185  C17AAG0186  C17AAG0187  C17AAG0188  C17AAG0189  C17AAG0190  C17AAG0191  C17AAG0192  C17AAG0193  C17AAG0194  C17AAG0195  C17AAG0197  C17AAG0198  C17AAG0199  C17AAG0200  C17AAG0201  C17AAG0202  C17AAG0203  C17AAG0204  C17AAG0205  C17AAG0206  C17AAG0207  C17AAG0208  C17AAG0209  C17AAG0210  C17AAG0211  C17AAG0212  C17AAG0213  C17AAG0215  C17AAG0216  C17AAG0217  C17AAG0219  C17AAG0220  C17AAG0221  C17AAG0222  C17AAG0223  C17AAG0224  C17AAG0225  C17AAG0226  C17AAG0227  C17AAG0228  C17AAG0230  C17AAG0231  C17AAG0232  C17AAG0233  C17AAG0234  C17AAG0235  C17AAG0236  C17AAG0237  C17AAG0238  C17AAG0239  C17AAG0240  C17AAG0241  C17AAG0242  C17AAG0243  C17AAG0244  C17AAG0245  C17AAG0247  C17AAG0248  C17AAG0249  C17AAG0250  C17AAG0251  C17AAG0252  C17AAG0253  C17AAG0254  C17AAG0255  C17AAG0256  C17AAG0257  C17AAG0258  C17AAG0259  C17AAG0260  C17AAG0261  C17AAG0262  C17AAG0263  C17AAG0264  C17AAG0265  C17AAG0266  C17AAG0267  C17AAG0268  C17AAG0269  C17AAG0271  C17AAG0272  C17AAG0273  C17AAG0274  C17AAG0275  C17AAG0276  C17AAG0277  C17AAG0278  C17AAG0280  C17AAG0281  C17AAG0282  C17AAG0283  C17AAG0284  C17AAG0285  C17AAG0286  C17AAG0288  C17AAG0290  C17AAG0291  C17AAG0292  C17AAG0293  C17AAG0294  C17AAG0295  C17AAG0297  C17AAG0298  C17AAG0299  C17AAG0300  C17AAG0301  C17AAG0302  C17AAG0303  C17AAG0304  C17AAG0305  C17AAG0306  C17AAG0307  C17AAG0308  C17AAG0309  C17AAG0310  C17AAG0311  C17AAG0312  C17AAG0313  C17AAG0315  C17AAG0317  C17AAG0318  C17AAG0319  C17AAG0320  C17AAG0321  C17AAG0322  C17AAG0323  C17AAG0324  C17AAG0325  C17AAG0326  C17AAG0327  C17AAG0328  C17AAG0329  C17AAG0330  C17AAG0331  C17AAG0332  C17AAG0333  C17AAG0334  C17AAG0335  C17AAG0336  C17AAG0337  C17AAG0338  C17AAG0339  C17AAG0340  C17AAG0341  C17AAG0342  C17AAG0343  C17AAG0344  C17AAG0345  C17AAG0346  C17AAG0347  C17AAG0348  C17AAG0349  C17AAG0350  C17AAG0351  C17AAG0352  C17AAG0353  C17AAG0354  C17AAG0355  C17AAG0356  C17AAG0357  C17AAG0358  C17AAG0359  C17AAG0360  C17AAG0361  C17AAG0362  C17AAG0363  C17AAG0364  C17AAG0365  C17AAG0366  C17AAG0367  C17AAG0368  C17AAG0369  C17AAG0370  C17AAG0371  C17AAG0372  C17AAG0373  C17AAG0374  C17AAG0375  C17AAG0376  C17AAG0377  C17AAG0378  C17AAG0379  C17AAG0380  C17AAG0381  C17AAG0386  C17AAG0387  C17AAG0388  C17AAG0389  C17AAG0390  C17AAG0391  C17AAG0392  C17AAG0393  C17AAG0394  C17AAG0395  C17AAG0396  C17AAG0397  C17AAG0398  C17AAG0399  C17AAG0400  C17AAG0401  C17AAG0402  C17AAG0404  C17AAG0405  C17AAG0406  C17AAG0407  C17AAG0408  C17AAG0409  C17AAG0410  C17AAG0411  C17AAG0414  C17AAG0415  C17AAG0416  C17AAG0417  C17AAG0419  C17AAG0421  C17AAG0422  C17AAG0423  C17AAG0424  C17AAG0425  C17AAG0426  C17AAG0427  C17AAG0428  C17AAG0429  C17AAG0430  C17AAG0432  C17AAG0433  C17AAG0434  C17AAG0435  C17AAG0436  C17AAG0437  C17AAG0438  C17AAG0439  C17AAG0440  C17AAG0441  C17AAG0442  C17AAG0444  C17AAG0446  C17AAG0447  C17AAG0448  C17AAG0449  C17AAG0450  C17AAG0451  C17AAG0452  C17AAG0453  C17AAG0454  C17AAG0455  C17AAG0456  C17AAG0457  C17AAG0458  C17AAG0459  C17AAG0460  C17AAG0461  C17AAG0462  C17AAG0463  C17AAG0464  C17AAG0465  C17AAG0466  C17AAG0467  C17AAG0468  C17AAG0469  C17AAG0470  C17AAG0471  C17AAG0472  C17AAG0473  C17AAG0474  C17AAG0475  C17AAG0476  C17AAG0477  C17AAG0478  C17AAG0479  C17AAG0480  C17AAG0481  C17AAG0482  C17AAG0486  C17AAG0487  C17AAG0488  C17AAG0489  C17AAG0490  C17AAG0492  C17AAG0493  C17AAG0494  C17AAG0495  C17AAG0496  C17AAG0497  C17AAG0498  C17AAG0499  C17AAG0500  C17AAG0501  C17AAG0502  C17AAG0503  C17AAG0504  C17AAG0505  C17AAG0506  C17AAG0507  C17AAG0508  C17AAG0509  C17AAG0510  C17AAG0511  C17AAG0512  C17AAG0513  C17AAG0514  C17AAG0515  C17AAG0516  C17AAG0517  C17AAG0518  C17AAG0519  C17AAG0520  C17AAG0521  C17AAG0522  C17AAG0523  C17AAG0524  C17AAG0525  C17AAG0526  C17AAG0527  C17AAG0528  C17AAG0529  C17AAG0530  C17AAG0531  C17AAG0532  C17AAG0533  C17AAG0534  C17AAG0536  C17AAG0537  C17AAG0538  C17AAG0539  C17AAG0540  C17AAG0541  C17AAG0543  C17AAG0544  C17AAG0545  C17AAG0546  C17AAG0547  C17AAG0549  C17AAG0550  C17AAG0551  C17AAG0552  C17AAG0553  C17AAG0554  C17AAG0555  C17AAG0557  C17AAG0558  C17AAG0559  C17AAG0560  C17AAG0561  C17AAG0562  C17AAG0563  C17AAG0564  C17AAG0565  C17AAG0566  C17AAG0567  C17AAG0568  C17AAG0569  C17AAG0570  C17AAG0571  C17AAG0572  C17AAG0573  C17AAG0574  C17AAG0575  C17AAG0576  C17AAG0577  C17AAG0578  C17AAG0579  C17AAG0580  C17AAG0582  C17AAG0583  C17AAG0584  C17AAG0585  C17AAG0586  C17AAG0587  C17AAG0588  C17AAG0589  C17AAG0590  C17AAG0591  C17AAG0592  C17AAG0593  C17AAG0594  C17AAG0595  C17AAG0596  C17AAG0597  C17AAH0001  C17AAH0002  C17AAH0005  C17AAH0007  C17AAH0008  C17AAH0009  C17AAH0010  C17AAH0011  C17AAH0012  C17AAH0041  C17AAH0042  C17AAH0043  C17AAH0044  C17AAH0045  C17AAH0046  C17AAH0047  C17AAH0048  C17AAH0049  C17AAH0051  C17AAH0052  C17AAH0053  C17AAH0054  C17AAH0055  C17AAH0056  C17AAH0057  C17AAH0058  C17AAH0059  C17AAH0060  C17AAH0061  C17AAH0062  C17AAH0063  C17AAH0064  C17AAH0065  C17AAH0066  C17AAH0067  C17AAH0068  C17AAH0069  C17AAH0071  C17AAH0072  C17AAH0074  C17AAH0076  C17AAH0077  C17AAH0078  C17AAH0079  C17AAH0080  C17AAH0081  C17AAH0082  C17AAH0083  C17AAH0084  C17AAH0085  C17AAH0142  C17AAH0143  C17AAH0144  C17AAH0146  C17AAH0147  C17AAH0148  C17AAH0149  C17AAH0152  C17AAH0153  C17AAH0154  C17AAH0155  C17AAH0156  C17AAH0157  C17AAH0159  C17AAH0160  C17AAH0161  C17AAH0162  C17AAH0164  C17AAH0174  C17AAH0179  C17AAH0181  C17AAH0182  C17AAH0183  C17AAH0184  C17AAH0191  C17AAH0193  C17AAH0196  C17AAH0197  C17AAH0198  C17AAH0199  C17AAH0200  C17AAH0201  C17AAH0202  C17AAH0203  C17AAH0204  C17AAH0205  C17AAH0206  C17AAH0207  C17AAH0208  C17AAH0209  C17AAH0210  C17AAH0211  C17AAH0212  C17AAH0213  C17AAH0214  C17AAH0215  C17AAH0217  C17AAH0218  C17AAH0219  C17AAH0220  C17AAH0221  C17AAH0285  C17AAH0286  C17AAH0287  C17AAH0289  C17AAH0290  C17AAH0291  C17AAH0292  C17AAH0294  C17AAH0296  C17AAH0297  C17AAH0298  C17AAH0299  C17AAH0300  C17AAH0301  C17AAH0302  C17AAH0303  C17AAH0304  C17AAH0305  C17AAH0306  C17AAH0307  C17AAH0309  C17AAH0310  C17AAH0311  C17AAH0312  C17AAH0313  C17AAH0314  C17AAH0315  C17AAH0316  C17AAH0318  C17AAH0319  C17AAH0320  C17AAH0321  C17AAH0322  C17AAH0323  C17AAH0324  C17AAH0325  C17AAH0326  C17AAH0328  C17AAH0329  C17AAH0330  C17AAH0331  C17AAH0332  C17AAH0333  C17AAH0334  C17AAH0335  C17AAH0336  C17AAH0337  C17AAH0338  C17AAH0339  C17AAH0341  C17AAH0342  C17AAH0344  C17AAH0345  C17AAH0346  C17AAH0349  C17AAH0351  C17AAH0352  C17AAH0353  C17AAH0354  C17AAH0355  C17AAH0356  C17AAH0357  C17AAH0358  C17AAH0360  C17AAH0361  C17AAH0362  C17AAH0365  C17AAH0366  C17AAH0368  C17AAH0369  C17AAH0370  C17AAH0371  C17AAH0372  C17AAH0374  C17AAH0375  C17AAH0376  C17AAH0379  C17AAH0380  C17AAH0381  C17AAJ0001  C17AAJ0002  C17AAJ0003  C17AAJ0004  C17AAJ0005  C17AAJ0006  C17AAJ0007  C17AAJ0008  C17AAJ0010  C17AAJ0011  C17AAJ0012  C17AAJ0014  C17AAJ0016  C17AAJ0017  C17AAJ0018  C17AAJ0019  C17AAJ0020  C17AAJ0022  C17AAJ0023  C17AAJ0024  C17AAJ0025  C17AAJ0026  C17AAJ0027  C17AAJ0028  C17AAJ0041  C17AAJ0042  C17AAJ0043  C17AAJ0044  C17AAJ0045  C17AAJ0048  C17AAJ0049  C17AAJ0050  C17AAJ0051  C17AAJ0052  C17AAJ0053  C17AAJ0054  C17AAJ0055  C17AAJ0056  C17AAJ0058  C17AAJ0059  C17AAJ0060  C17AAJ0063  C17AAJ0064  C17AAJ0065  C17AAJ0066  C17AAJ0067  C17AAJ0068  C17AAJ0070  C17AAJ0072  C17AAJ0076  C17AAJ0077  C17AAJ0078  C17AAJ0079  C17AAJ0080  C17AAJ0083  C17AAJ0089  C17AAJ0090  C17AAJ0091  C17AAJ0093  C17AAJ0094  C17AAJ0100  C17AAJ0103  C17AAJ0104  C17AAJ0105  C17AAJ0109  C17AAJ0114  C17AAJ0115  C17AAJ0116  C17AAJ0122  C17AAJ0123  C17AAJ0124  C17AAJ0125  C17AAJ0126  C17AAJ0128  C17AAJ0129  C17AAJ0130  C17AAJ0131  C17AAJ0133  C17AAJ0134  C17AAJ0139  C17AAJ0141  C17AAJ0142  C17AAJ0144  C17AAJ0145  C17AAJ0146  C17AAJ0148  C17AAJ0149  C17AAJ0150  C17AAJ0151  C17AAJ0152  C17AAJ0153  C17AAJ0154  C17AAJ0155  C17AAJ0156  C17AAJ0157  C17AAJ0158  C17AAJ0159  C17AAJ0160  C17AAJ0164  C17AAJ0167  C17AAJ0168  C17AAJ0169  C17AAJ0171  C17AAJ0172  C17AAJ0175  C17AAJ0176  C17AAJ0177  C17AAJ0181  C17AAJ0182  C17AAJ0184  C17AAJ0186  C17AAJ0187  C17AAJ0189  C17AAJ0190  C17AAJ0192  C17AAJ0193  C17AAJ0196  C17AAJ0198  C17AAJ0199  C17AAJ0202  C17AAJ0205  C17AAJ0206  C17AAJ0208  C17AAJ0209  C17AAJ0210  C17AAJ0212  C17AAJ0213  C17AAJ0214  C17AAJ0219  C17AAJ0223  C17AAJ0226  C17AAJ0227  C17AAJ0228  C17AAJ0229  C17AAJ0230  C17AAJ0231  C17AAJ0232  C17AAJ0235  C17AAJ0238  C17AAJ0239  C17AAJ0281  C17AAJ0282  C17AAJ0283  C17AAJ0284  C17AAJ0286  C17AAJ0287  C17AAJ0291  C17AAJ0297  C17AAJ0298  C17AAJ0299  C17AAJ0301  C17AAJ0302  C17AAJ0304  C17AAJ0305  C17AAJ0308  C17AAJ0309  C17AAJ0310  C17AAJ0314  C17AAJ0315  C17AAK0011  C17AAK0012  C17AAK0013  C17AAK0014  C17AAK0015  C17AAK0016  C17AAK0017  C17AAK0019  C17AAK0020  C17AAK0023  C17AAK0024  C17AAK0025  C17AAK0026  C17AAK0027  C17AAK0028  C17AAK0030  C17AAK0031  C17AAK0033  C17AAK0034  C17AAK0035  C17AAK0036  C17AAK0037  C17AAK0039  C17AAK0040  C17AAK0081  C17AAK0083  C17AAK0084  C17AAK0086  C17AAK0087  C17AAK0088  C17AAK0089  C17AAK0092  C17AAK0093  C17AAK0094  C17AAK0095  C17AAK0097  C17AAK0100  C17AAK0111  C17AAK0114  C17AAK0116  C17AAK0120  C17AAK0125  C17AAK0126  C17AAK0129  C17AAK0132  C17AAK0133  C17AAK0161  C17AAK0163  C17AAK0164  C17AAK0174  C17AAK0218",
      "center_classification_date": "20180327",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide).",
      "state": "CT",
      "product_description": "Signia\" Power Handle (SIGPHANDLE, SIGNVAL1 and SIGVAL20 - Kits)    The Signia\" Power Handle, when used with Endo GIA\" single-use reloads and Endo GIA\" single-use reloads with Tri-Staple\" Technology, has applications in open and minimally invasive general abdominal, gynecological, pediatric, and thoracic surgery for resection, transection, and creation of anastomosis.",
      "report_date": "20180404",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Covidien LLC",
      "recall_number": "Z-1237-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79483",
      "termination_date": "20200414",
      "more_code_info": "",
      "recall_initiation_date": "20171213",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}