      "status": "Terminated",
      "city": "Boca Raton",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91795",
      "recalling_firm": "Sensus Healthcare, Inc.",
      "address_1": "851 Broken Sound Pkwy Nw Ste 215",
      "address_2": "",
      "postal_code": "33487-3635",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Distribution to states of:  RX, IN, NM, FL, NY, FA, NC, IA, and IL.",
      "recall_number": "Z-1235-2023",
      "product_description": "Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities",
      "product_quantity": "23 units",
      "reason_for_recall": "When the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.",
      "recall_initiation_date": "20230210",
      "center_classification_date": "20230308",
      "termination_date": "20240131",
      "report_date": "20230315",
      "code_info": "Software version 1.9.2 loaded onto Serial Numbers: 1512-2002; 1512-2003; 1606-2007; 1609-2010; 1609-2011; 1609-2018; 1612-2023; 1612-2025; 1705-2032; 1709-2046; 1803-2068; 1906-2137; 1909-2149; 1912-2167; 2003-2177; 2112-2201; 2112-2202; 2112-2208; 2112-2218; 2203-2232; 2203-2241; 2206-2264; 2206-2272",
      "more_code_info": ""
    }
  ]
}