  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.",
      "address_2": "",
      "product_quantity": "3454",
      "code_info": "915209320  915209340  915209360  915209380  915209400  915209420  915209440  915209460  915211320  915211340  915211360  915211380  915211400  915211420  915211440  915211460  915213320  915213340  915213360  915213380  915213400  915213420  915213440  915213460  916209320  916209340  916209360  916209380  916209400  916209420  916209440  916209460  916211320  916211340  916211360  916211380  916211400  916211420  916211440  916211460  916213320  916213340  916213360  916213380  916213400  916213420  916213440  916213460  925200320  925209320  925209340  925209360  925209380  925209400  925209420  925209440  925209460  925211340  925211360  925211380  925211400  925211420  925211440  925211460  925213320  925213340  925213360  925213380  925213400  925213420  925213440  925213460  926209320  926209340  926209360  926209380  926209400  926209420  926209440  926209460  926211320  926211340  926211360  926211380  926211400  926211420  926211440  926213320  926213340  926213360  926213380  926213400  926213420  926213440  926213460  935209320  935209340  935209360  935209380  935209400  935209420  935209440  935209460  935211320  935211340  935211360  935211380  935211400  935211420  935211440  935211460  935213320  935213340  935213360  935213380  935213400  935213420  935213440  935213460  936213320  936213340  936213360  936213380  936213400  936213420  936213440  936213460  936309320  936309340  936309360  936309380  936309400  936309420  936309440  936309460  936311320  936311340  936311360  936311380  936311400  936311420  936311440  936311460",
      "center_classification_date": "20180327",
      "distribution_pattern": "USA (nationwide)",
      "state": "IN",
      "product_description": "ACE Trochanteric Nail System  STER L H TROC NAIL 320X9X125  STER L H TROC NAIL 340X9X125  STER L H TROC NAIL 360X9X125  STER L H TROC NAIL 380X9X125  STER L H TROC NAIL 400X9X125  STER L H TROC NAIL 420X9X125  STER L H TROC NAIL 440X9X125  STER L H TROC NAIL 460X9X125  STER L H TROC NAIL 320X11X125  STER L H TROC NAIL 340X11X125  STER L H TROC NAIL 360X11X125  STER L H TROC NAIL 380X11X125  STER L H TROC NAIL 400X11X125  STER L H TROC NAIL 420X11X125  STER L H TROC NAIL 440X11X125  STER L H TROC NAIL 460X11X125  STER L H TROC NAIL 320X13X125  STER L H TROC NAIL 340X13X125  STER L H TROC NAIL 360X13X125  STER L H TROC NAIL 380X13X125  STER L H TROC NAIL 400X13X125  STER L H TROC NAIL 420X13X125  STER L H TROC NAIL 440X13X125  STER L H TROC NAIL 460X13X125  STER R H TROC NAIL 320X9X125  STER R H TROC NAIL 340X9X125  STER R H TROC NAIL 360X9X125  STER R H TROC NAIL 380X9X125  STER R H TROC NAIL 400X9X125  STER R H TROC NAIL 420X9X125  STER R H TROC NAIL 440X9X125  STER R H TROC NAIL 460X9X125  STER R H TROC NAIL 320X11X125  STER R H TROC NAIL 340X11X125  STER R H TROC NAIL 360X11X125  STER R H TROC NAIL 380X11X125  STER R H TROC NAIL 400X11X125  STER R H TROC NAIL 420X11X125  STER R H TROC NAIL 440X11X125  STER R H TROC NAIL 460X11X125  STER R H TROC NAIL 320X13X125  STER R H TROC NAIL 340X13X125  STER R H TROC NAIL 360X13X125  STER R H TROC NAIL 380X13X125  STER R H TROC NAIL 400X13X125  STER R H TROC NAIL 420X13X125  STER R H TROC NAIL 440X13X125  STER R H TROC NAIL 460X13X125  STER L H TROC NAIL 320X11X130  STER L H TROC NAIL 320X9X130  STER L H TROC NAIL 340X9X130  STER L H TROC NAIL 360X9X130  STER L H TROC NAIL 380X9X130  STER L H TROC NAIL 400X9X130  STER L H TROC NAIL 420X9X130  STER L H TROC NAIL 440X9X130  STER L H TROC NAIL 460X9X130  STER L H TROC NAIL 340X11X130  STER L H TROC NAIL 360X11X130  STER L H TROC NAIL 380X11X130  STER L H TROC NAIL 400X11X130  STER L H TROC NAIL 420X11X130  STER L H TROC NAIL 440X11X130  STER L H TROC NAIL 460X11X130  STER L H TROC NAIL 320X13X130  STER L H TROC NAIL 340X13X130  STER L H TROC NAIL 360X13X130  STER L H TROC NAIL 380X13X130  STER L H TROC NAIL 400X13X130  STER L H TROC NAIL 420X13X130  STER L H TROC NAIL 440X13X130  STER L H TROC NAIL 460X13X130  STER R H TROC NAIL 320X9X130  STER R H TROC NAIL 340X9X130  STER R H TROC NAIL 360X9X130  STER R H TROC NAIL 380X9X130  STER R H TROC NAIL 400X9X130  STER R H TROC NAIL 420X9X130  STER R H TROC NAIL 440X9X130  STER R H TROC NAIL 460X9X130  STER R H TROC NAIL 320X11X130  STER R H TROC NAIL 340X11X130  STER R H TROC NAIL 360X11X130  STER R H TROC NAIL 380X11X130  STER R H TROC NAIL 400X11X130  STER R H TROC NAIL 420X11X130  STER R H TROC NAIL 440X11X130  STER R H TROC NAIL 320X13X130  STER R H TROC NAIL 340X13X130  STER R H TROC NAIL 360X13X130  STER R H TROC NAIL 380X13X130  STER R H TROC NAIL 400X13X130  STER R H TROC NAIL 420X13X130  STER R H TROC NAIL 440X13X130  STER R H TROC NAIL 460X13X130  STER L H TROC NAIL 320X9X135  STER L H TROC NAIL 340X9X135  STER L H TROC NAIL 360X9X135  STER L H TROC NAIL 380X9X135  STER L H TROC NAIL 400X9X135  STER L H TROC NAIL 420X9X135  STER L H TROC NAIL 440X9X135  STER L H TROC NAIL 460X9X135  STER L H TROC NAIL 320X11X135  STER L H TROC NAIL 340X11X135  STER L H TROC NAIL 360X11X135  STER L H TROC NAIL 380X11X135  STER L H TROC NAIL 400X11X135  STER L H TROC NAIL 420X11X135  STER L H TROC NAIL 440X11X135  STER L H TROC NAIL 460X11X135  STER L H TROC NAIL 320X13X135  STER L H TROC NAIL 340X13X135  STER L H TROC NAIL 360X13X135  STER L H TROC NAIL 380X13X135  STER L H TROC NAIL 400X13X135  STER L H TROC NAIL 420X13X135  STER L H TROC NAIL 440X13X135  STER L H TROC NAIL 460X13X135  STER R H TROC NAIL 320X13X135  STER R H TROC NAIL 340X13X135  STER R H TROC NAIL 360X13X135  STER R H TROC NAIL 380X13X135  STER R H TROC NAIL 400X13X135  STER R H TROC NAIL 420X13X135  STER R H TROC NAIL 440X13X135  STER R H TROC NAIL 460X13X135  STER R H TROC NAIL 320X9X135  STER R H TROC NAIL 340X9X135  STER R H TROC",
      "report_date": "20180404",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-1235-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79408",
      "termination_date": "20200422",
      "more_code_info": "",
      "recall_initiation_date": "20171220",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}