  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Saint Charles",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91773",
      "recalling_firm": "Elekta Inc",
      "address_1": "1450 Beale St Ste 205",
      "address_2": "",
      "postal_code": "63303",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Nationwide distribution in the US.  International distribution to Albania, Algeria, Antigua and Barbuda, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegowina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Lao People's Democratic Republic, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Macedonia, Malaysia, Malta, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestinian Territory, Occupied, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Zimbabwe.",
      "recall_number": "Z-1233-2023",
      "product_description": "Monaco RTP System.      Used to make treatment plans for patients with prescriptions for external beam radiation therapy.",
      "product_quantity": "2,020 units",
      "reason_for_recall": "Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation.",
      "recall_initiation_date": "20230228",
      "center_classification_date": "20230307",
      "report_date": "20230315",
      "code_info": "Software Build: 5.11.00, 5.11.01, 5.11.02, 5.11.03;    UDI Numbers: (01)00858164002169(10)5.11.00, (01)00858164002169(10)5.11.01, (01)00858164002169(10)5.11.02, (01)00858164002169(10)5.11.03",
      "more_code_info": ""
    }
  ]
}