      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64682",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.",
      "recall_number": "Z-1233-2013",
      "product_description": "Stryker Irrigator Reusable Tips:  5mm X 45cm Regular Tip, vented, Non-Reflective Finish.    Suction irrigator tips aid in cleaning and clearing the operative site during  laparoscopic surgery.",
      "product_quantity": "147,921 - all sizes",
      "reason_for_recall": "The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.",
      "recall_initiation_date": "20130327",
      "center_classification_date": "20130503",
      "termination_date": "20130924",
      "report_date": "20130515",
      "code_info": "Manufacturer Part Number: 250-070-488"
    }
  ]
}