{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91642",
      "recalling_firm": "Datascope Corp.",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide US distribution. Worldwide international distribution.",
      "recall_number": "Z-1232-2023",
      "product_description": "Cardiosave Rescue. Model Number: 0998-00-0800-83.",
      "product_quantity": "47 units (44 US, 3 OUS)",
      "reason_for_recall": "Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.",
      "recall_initiation_date": "20230207",
      "center_classification_date": "20230306",
      "report_date": "20230315",
      "code_info": "Model Number: 0998-00-0800-83. UDI-DI: 10607567108407. All serial numbers.",
      "more_code_info": ""
    }
  ]
}