  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Elkton",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76346",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "125 Blue Ball Rd",
      "address_2": "N/A",
      "postal_code": "21921-5315",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "The product was distributed domestically and also to foreign consignees.  But not to government agency's or Canada.",
      "recall_number": "Z-1232-2017",
      "product_description": "Titan\" Stabilizer Attachment and Titan TM  360 Stabilizer  Attachment",
      "product_quantity": "1463",
      "reason_for_recall": "Terumo Cardiovascular  Systems (Terumo CVS) is voluntarily recalling certain lots of Titan\" Stabilizers and Titan\" 360 Stabilizers because the Suction Tubing Clip may break when it  is being attached to a Terumo CVS Hercules Stabilizing Arm.  Although the clip is for convenience purposes and is not required for device operation, there is a potential for broken Suction Tubing Clip fragments to fall into the surgical site.",
      "recall_initiation_date": "20150105",
      "center_classification_date": "20170222",
      "termination_date": "20170223",
      "report_date": "20170301",
      "code_info": "Catalog Number: 401-231, Product Description: Titan\" Stabilizer Attachment , Lot Number: 052R      Catalog Number 401-231u , Product Description TitanTM  360 Stabilizer  Attachment , Lot Number: 020R, 024R, 026R, 029R, 030R,034R,035R,037R,  040 , 048R"
    }
  ]
}