  "results": [
    {
      "status": "Ongoing",
      "city": "West Chester",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91618",
      "recalling_firm": "Aniara Diagnostica LLC",
      "address_1": "7768 Service Center Dr",
      "address_2": "",
      "postal_code": "45069-2442",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the state of NC.",
      "recall_number": "Z-1227-2023",
      "product_description": "BIOPHEN Protein C 5 kit is for in vitro quantitative determination of Protein C activity on human citrated plasma. Protein C is a glycoprotein, vitamin K dependent, which inhibits coagulation. Protein C in plasma is measured following a specific activation with Protac (Agkistrodom Contortrix), an enzyme extracted from snake venom. Activated protein C hydrolysis the chromogenic substrate (SaPC-21) which release para-nitroaniline (pNA). The amount of pNA released (measured by absorbance at 405 nm) is directly proportional to the concentration of Protein C in the specimen.",
      "product_quantity": "2 units",
      "reason_for_recall": "New warnings and precautionary statements on product labelling: Reagent 2 (R2) of BIOPHEN Protein C 5 contains the chemical substance - Cesium Chloride which changed classification and has become hazardous according to the CLP regulation ((EC) No 1272/2008) is suspected of damaging fertility",
      "recall_initiation_date": "20230201",
      "center_classification_date": "20230306",
      "report_date": "20230315",
      "code_info": "UDI-DI: 366353700063BD Lot FA083327 (And all future lots)",
      "more_code_info": ""
    }
  ]
}