  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90076",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) to AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and OUS (internationally) to Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Columbia, Cote D'Ivoire, Mexico, Uruguay, United Kingdom, United Arab Emirates, Turkey, Thailand, Tanzania, Sweden, Spain, Singapore, Saudi Arabia, Russia, Romania, Portugal, Poland, Philippines, Panama, Omar, Norway, Nigeria, Nicaragua, New Zealand, Netherlands, Malaysia, Latvia, Kuwait, Korea, Kenya, Kazakhstan, Japan, Italy, Ireland, Indonesia, India, Hungary, Germany, France, Finland, Estonia, Egypt, Ecuador, Denmark, Czech Republic.",
      "recall_number": "Z-1227-2022",
      "product_description": "Replacement back up batteries distributed on or after April 1, 2019, for CARESCAPE R860, Engstr¿om Carestation and Engstr¿om PRO Ventilators.",
      "product_quantity": "88,379 batteries",
      "reason_for_recall": "Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.",
      "recall_initiation_date": "20220418",
      "center_classification_date": "20220606",
      "report_date": "20220615",
      "code_info": "Field Replacement Unit kits PN: 1009-5682-000-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V) and PN: 5856787-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V PAIR).",
      "more_code_info": ""
    }
  ]
}