{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Austin",
      "address_1": "7000 Burleson Rd Bldg D",
      "reason_for_recall": "Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.",
      "address_2": "",
      "product_quantity": "260",
      "code_info": "Lots: N18446, N18640-1, R19208, R19406",
      "center_classification_date": "20190604",
      "distribution_pattern": "US: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM.    Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Poland, South Africa, Chile",
      "state": "TX",
      "product_description": "ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV16F",
      "report_date": "20190612",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "OriGen Biomedical, Inc.",
      "recall_number": "Z-1227-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "82512",
      "termination_date": "20210727",
      "more_code_info": "",
      "recall_initiation_date": "20170117",
      "postal_code": "78744-3202",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}