  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64682",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.",
      "recall_number": "Z-1226-2013",
      "product_description": "Stryker Irrigator Reusable Tips:  2mm X 32cm Regular Tip with 1 Hole, Metal Finish.    Suction irrigator tips aid in cleaning and clearing the operative site during  laparoscopic surgery.",
      "product_quantity": "147,921 - all sizes",
      "reason_for_recall": "The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.",
      "recall_initiation_date": "20130327",
      "center_classification_date": "20130503",
      "termination_date": "20130924",
      "report_date": "20130515",
      "code_info": "Manufacturer Part Number: 250-070-471"
    }
  ]
}