      "status": "Terminated",
      "city": "Arlington",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90140",
      "recalling_firm": "MicroPort Orthopedics Inc.",
      "address_1": "5677 Airline Rd",
      "address_2": "N/A",
      "postal_code": "38002-9501",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of FL, IN, MO, CA and the country of Canada.",
      "recall_number": "Z-1225-2022",
      "product_description": "PROFEMUR PLASMA Z CLASSIC STEM SIZE 7 STANDARD SHORT NECK",
      "product_quantity": "8 units",
      "reason_for_recall": "One manufacturing lot is missing the plasma-spray coating in the proximal region of the stem body.",
      "recall_initiation_date": "20220504",
      "center_classification_date": "20220602",
      "termination_date": "20240116",
      "report_date": "20220608",
      "code_info": "Catalog Number(s): PHAPCLS7; Lot Number(s): 1903727; Expiration Date: 11/24/2029; UDI/DI Number: 00192629112773"
    }
  ]
}