      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82366",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of: AZ, CA, FL, IL, OH, TX, and VA.    The products were distributed to the following foreign countries:  Australia, Belgium, Canada, Chile, Columbia, France, India, Italy, Japan, Portugal, Singapore, Spain, United Kingdom",
      "recall_number": "Z-1222-2019",
      "product_description": "VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658    Product Usage:  The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents",
      "product_quantity": "80",
      "reason_for_recall": "One of the software algorithms used to detect sample dispense errors was inadvertently disabled.  Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the user.",
      "recall_initiation_date": "20190311",
      "center_classification_date": "20190425",
      "termination_date": "20201207",
      "report_date": "20190501",
      "code_info": "Systems Running Software Version 3.4 & 3.4.1"
    }
  ]
}