      "country": "United States",
      "city": "San Diego",
      "address_1": "9900 Old Grove Rd",
      "reason_for_recall": "The device power supply was not manufactured according to specification which can result in the loss of the protective grounding of the instrument.",
      "address_2": "",
      "product_quantity": "1 instrument",
      "code_info": "Serial number 4314",
      "center_classification_date": "20180327",
      "distribution_pattern": "There was no U.S. distribution.  Distribution was made to Panama and United Kingdom.",
      "state": "CA",
      "product_description": "QUANTA-Lyser 240 IFA, Part #GS0242.    The instrument is a fully automated, high-throughput, bench top system designed for clinical use to automate slide and reagent barcode reading, sample dilution and distribution, reagent pipetting and dispensing, slide and plate washing, dispensing slide mounting medium in the preparation for microscope analysis, and photometric measurement of microwell plates.",
      "report_date": "20180404",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Inova Diagnostics Incorporated",
      "recall_number": "Z-1221-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79380",
      "termination_date": "20180522",
      "more_code_info": "",
      "recall_initiation_date": "20180126",
      "postal_code": "92131-1638",
      "voluntary_mandated": "Voluntary: Firm initiated",