{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Covington",
      "address_1": "8195 Industrial Blvd NE",
      "reason_for_recall": "Leaks or disconnect near the sampling port of the Intra Abdominal Pressure Monitoring Device which could lead to the potential of inaccurate readings.  The measurement should be confirmed with other means, such as a new IAP system of the same kind or by choosing other diagnostic means such as but not limited to alternative systems, such as open IAP measurement options or exchanging the tubing system, which does not require invasive maneuvers.",
      "address_2": "",
      "product_quantity": "1,300 total",
      "code_info": "Product Code IAP001 lot numbers NGDQ2943 and NGDS0622.",
      "center_classification_date": "20200213",
      "distribution_pattern": "Distributed Nationwide including Washington DC and Guaynabo PR; Distributed OUS to Belgium and Japan.",
      "state": "GA",
      "product_description": "The Bard Intra-abdominal Pressure Monitoring Device is composed of a tubing set used for infusing fluid into the urinary bladder through the Foley catheter sampling port. It utilizes a clamping device to occlude the urinary drainage tubing to form a fluid column through which pressure is measured.",
      "report_date": "20200219",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "C.R. Bard Inc",
      "recall_number": "Z-1219-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "84744",
      "termination_date": "20210412",
      "more_code_info": "",
      "recall_initiation_date": "20200113",
      "postal_code": "30014-1497",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}