  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76178",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: 0    Foreign: Singapore & Malaysia",
      "recall_number": "Z-1219-2017",
      "product_description": "Vanguard Total Knee, PUNCH THRU TRL PLATES, 63 MM Item No: 32-487261, 67 MM Item No: 32-487262, 71 MM Item No: 32-487263, 75 MM Item No: 32-487264, and 79 MM Item No: 32-487265",
      "product_quantity": "26",
      "reason_for_recall": "Zimmer Biomet is conducting a medical device recall for ARCOS & TPRLC broaches and rasps due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.",
      "recall_initiation_date": "20161215",
      "center_classification_date": "20170216",
      "termination_date": "20180629",
      "report_date": "20170222",
      "code_info": "Vanguard Total Knee, PUNCH THRU TRL PLATES, 63 MM Item No: 32-487261, 67 MM Item No: 32-487262, 71 MM Item No: 32-487263, 75 MM Item No: 32-487264, and 79 MM Item No: 32-487265; Lot No's: ZB160701, ZB160702, ZB160703, ZB160701, ZB160702, ZB160701, ZB160702, ZB160703, ZB160801, ZB160802, ZB160803, ZB160801, ZB160802",
      "more_code_info": ""
    }
  ]
}