{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80887",
      "recalling_firm": "Smiths Medical ASD Inc.",
      "address_1": "6000 Nathan Ln N",
      "address_2": "N/A",
      "postal_code": "55442-1690",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Canada, Australia, and New Zealand.",
      "recall_number": "Z-1218-2019",
      "product_description": "Jelco(R) ViaValve(TM) safety intravenous (I.V.) catheters, 18G x 1 1/4\", Model Number REF 3265, MOD 10     The Jelco(R) ViaValve(TM) safety I.V. catheter is a sterile, non pyrogenic, single use, radiopaque intravascular catheter device used for the administration of medically prescribed fluids, designed to reduce the risk of accidental secondary needle sticks.",
      "product_quantity": "136,600 units",
      "reason_for_recall": "Several lots of Jelco¿ ViaValve\" safety intravenous (I.V.) catheters exhibited a V shaped puncture at the base of the catheter near the hub due to an adjustment that was made in the manufacturing process.",
      "recall_initiation_date": "20180822",
      "center_classification_date": "20190425",
      "termination_date": "20210211",
      "report_date": "20190501",
      "code_info": "Lot Numbers: 3584467, 3584468, 3584469, 3576073, 3576233, 3578511, 3578512"
    }
  ]
}