  "results": [
    {
      "country": "United States",
      "city": "Gainesville",
      "address_1": "2320 NW 66th Ct",
      "reason_for_recall": "The products have been identified to contain springs which have been manufactured with type-304 stainless steel rather than the specified titanium alloy.",
      "address_2": "",
      "product_quantity": "242",
      "code_info": "size 3.5mm x 10mm, Catalog #05-000-20-3510, Lot Number #35917001.  size 3.5mm x 14mm, Catalog #05-000-20-3514, Lot Number 35917002 and 40023002.  size 3.5mm x 20mm, Catalog #05-000-20-3520, Lot Number 35917003.  size 3.5mm x 26mm, Catalog #05-000-20-3526, Lot Number 35917004.  size 4.0mm x 14mm, Catalog #05-000-20-4014, Lot Number 35917005.  size 4.0mm x  20mm, Catalog #05-000-20-4020, Lot Number 35917006.  size 4.0mm x 26mm, Catalog #05-000-20-4026, Lot Number 35917007.  1/4 Shank, size 3.5mm x 26mm, Catalog #05-000-20-3526, Lot Number 35917008.  1/4 Shank, size 3.5mm x 30mm, Catalog #05-000-20-3530, Lot Number 35917009.  1/4 Shank, size 4.0mm x 26mm, Catalog #05-000-20-4026, Lot Number 35917010.  1/4 Shank, size 4.0mm x 30mm, Catalog #05-000-20-4030, Lot Number 35917011.",
      "center_classification_date": "20140318",
      "distribution_pattern": "USA Distribution including the states of  FL, NY, OR, IL, PA, MO and NJ.",
      "state": "FL",
      "product_description": "Gibralt Spine System Polyaxial Screw    Intended to promote fusion of the cervical spine, and the thoracic spine, (C3-T3), the Gibralt Spinal System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C3-T3) spine.",
      "report_date": "20140326",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Exactech, Inc.",
      "recall_number": "Z-1217-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67486",
      "termination_date": "20140804",
      "more_code_info": "",
      "recall_initiation_date": "20140106",
      "postal_code": "32653-1630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}