  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Saint Paul",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91646",
      "recalling_firm": "Boston Scientific Corporation",
      "address_1": "4100 Hamline Ave N",
      "address_2": "",
      "postal_code": "55112-5700",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution to the healthcare facilities and healthcare providers are located in CA, CO, DE, FL, GA, ID, IL, IN, IA, KY, ME, MD, MA, NV, NH, NJ, NY, NC, OH, PA, RI, SC, TN TX, VA, WA, WV, and WI.",
      "recall_number": "Z-1216-2023",
      "product_description": "LATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460.",
      "product_quantity": "1,531 patient data records",
      "reason_for_recall": "Under specific circumstances, the U.S. product registration system did not send up enablement requests to U.S. LATITUDE Customer Support during an 8-month interval in 2022.  For these patients, the HeartLogic Index and any associated Yellow Alerts are currently not visible in LATITUDE as may have been intended by the healthcare provider..",
      "recall_initiation_date": "20230202",
      "center_classification_date": "20230303",
      "report_date": "20230315",
      "code_info": "UDI-DI: 00802526613876",
      "more_code_info": ""
    }
  ]
}