{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91612",
      "recalling_firm": "Baxter Healthcare Corporation",
      "address_1": "1 Baxter Pkwy",
      "address_2": "",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution and Canada.",
      "recall_number": "Z-1213-2023",
      "product_description": "Baxter MiniCap Disconnect Cap with Povidone-Iodine Solution, 5C4466P; Accessory to disposable administration set for peritoneal dialysis",
      "product_quantity": "33,866,400 units",
      "reason_for_recall": "MiniCap Disconnect Cap with Povidone-Iodine are packaged in foil pouches, which may have been incorrectly sealed, i.e., the pouches may have open or weak seals. This could lead to exposure to air, resulting in insufficient iodine/dry sponge inside the MiniCap, which could lead to the potential for inadequate disinfectant.",
      "recall_initiation_date": "20230201",
      "center_classification_date": "20230303",
      "report_date": "20230315",
      "code_info": "UDI/DI 00085412007694, Lot numbers:  GD912051, exp 31-Mar-2024; GD912068, exp 30-Apr-2024; GD912099, exp 30-Apr-2024; GD912112, exp 30-Apr-2024; GD912143, exp 30-Apr-2024; GD912204, exp 31-May-2024.    Expansion on 05/10/2023: Product Code 5C4466P with all lots numbers manufactured prior to 09/30/2022.",
      "more_code_info": ""
    }
  ]
}