      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64918",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including DC and Puerto Rico, excluding the state of RI and the countries of Yugoslavia YEMEN Viet Nam Venezuela Uzbekistan Uruguay United Arab Emirates Tunisia Thailand Tanzania Taiwan Turkey Syria Switzerland Sweden Spain South Korea South Africa Slovenia Slovakia Singapore Saudi Arabia Russia Romania Reunion Qatar Portugal Poland Philippines Peru Pakistan Paraguay Philippines Oman Norway Nigeria New Zealand Netherlands Morocco Mexico Martinique Malaysia Macedonia Mauritania Luxembourg Lithuania Libya Lebanon Latvia Kuwait Kenya Korea Korea (Republic Of [South] Korea) KUWAIT Kazakhstan Jordan Japan Jamaica Italy Israel Ireland Iran Iraq Indonesia India Hungary Hong Kong Honduras Guadeloupe Greece Germany Georgia, Republic of French Polynesia France Finland Estonia Ecuador Egypt Denmark Dominican Republic Czech Republic Croatia Colombia China Chile Canada Bulgaria Brazil Bosnia And Herzegovina Belgium Belarus Barbados Bahrain Austria Australia Argentina Arab Emirates Algeria and United Kingdom.",
      "recall_number": "Z-1208-2013",
      "product_description": "GE Healthcare, GE 1.5T SIGNA HDe MR SYSTEM.    Indicated for use as a diagnostic imaging device.",
      "product_quantity": "822 (154 US; 668 OUS)",
      "reason_for_recall": "GE Healthcare has recently become aware of a Reformat Image Flip Issue associated with GE MR Products.    The axial and coronal reformatted images from a sagittal 3D series (such as with applications of FSPGR-3D T1, BRAVO T1, or CUBE T1) of over 126 slices may display a left-to-right image flip .  In addition, the acquired sagittal images may be annotated with left-right location annotation revers",
      "recall_initiation_date": "20120608",
      "center_classification_date": "20130501",
      "termination_date": "20141103",
      "report_date": "20130508",
      "code_info": "Mfg Lot or Serial #  00000004984YR1  00000005068YR2  00000005079YR9  00000005080YR7  00000005081YR5  00000005648YR1  00000005656YR4  00000005659YR8  00000005664YR8  00000006155YR6  00000006157YR2  00000006172YR1  00000006176YR2  00000006177YR0  00000006188YR7  00000006194YR5  00000006198YR6  00000006200YR0  00000006204YR2  00000006514YR4  00000006515YR1  00000006516YR9  00000006519YR3  00000006522YR7  00000006525YR0  00000006528YR4  00000006532YR6  00000006535YR9  00000006538YR3  00000006544YR1  00000007947YR5  00000007954YR1  00000007969YR9  00000007970YR7  00000007971YR5  00000008692YR6  00000008698YR3  00000008699YR1  00000008701YR5  00000008703YR1  00000008709YR8  00000008715YR5  00000008716YR3  00000008720YR5  00000008726YR2  00000008727YR0  00000008728YR8  00000008729YR6  00000008732YR0  00000008734YR6  00000009833YR5  00000010024YR8  00000010025YR5  00000010114YR7  00000010578YR3  00000010580YR9  00000010583YR3  00000010686YR4  00000010794YR6  00000010795YR3  00000010830YR8  00000010831YR6  00000010834YR0  00000010856YR3  00000010863YR9  00000011210YR2  00000011428YR0  00000011429YR8  00000011495YR9  00000011499YR1  00000011637YR6  00000011648YR3  00000011650YR9  00000011669YR9  00000011686YR3  00000011826YR5  00000011829YR9  00000011830YR7  00000011842YR2  00000012085YR7  00000012098YR0  00000012099YR8  00000012100YR4  00000012102YR0  00000012142YR6  00000012172YR3  00000012206YR9  00000012207YR7  00000012221YR8  00000012227YR5  00000012286YR1  00000012287YR9  00000012289YR5  00000012330YR7  00000012393YR5  00000012434YR7  00000012435YR4  00000012503YR9  00000012507YR0  00000012576YR5  00000012651YR6  00000012652YR4  00000012727YR4  00000012728YR2  00000012755YR5  00000012756YR3  00000012818YR1  00000012820YR7  00000012940YR3  00000012943YR7  00000012988YR2  00000012992YR4  00000012994YR0  00000012995YR7  00000013056YR7  00000013061YR7  00000013177YR1  00000013180YR5  00000013185YR4  00000013239YR9  00000013241YR5  00000013246YR4  00000013301YR7  00000013306YR6  00000013310YR8  00000013375YR1  00000013379YR3  00000013384YR3  00000013385YR0  00000013399YR1  00000013453YR6  00000013799YR2  00000013800YR8  00000013802YR4  00000013804YR0  00000013878YR4  00000013880YR0  00000014113YR5  00000014114YR3  00000014214YR1  00000014276YR0  00000014277YR8  00000014520YR1  00000014524YR3  00000014801YR5  00000014806YR4  00000014858YR5  00000014859YR3  00000014907YR0  00000014909YR6  00000014910YR4  00000014956YR7  00000015146YR4  00000013380YR1  00000011535YR2  00000012209YR3  00000014414YR7  00000006174YR7  00000006507YR8  00000010827YR4  00000012144YR2  00000012497YR4  00000012575YR7  00000012753YR0  00000012991YR6  00000013693YR7  00000013875YR0  00000014415YR4  00000014419YR6  00000014620YR9  00000011692YR1  00000012367YR9  00000012508YR8  00000012550YR0  00000012722YR5  00000012942YR9  00000013305YR8  00000013451YR0  00000013626YR7  00000013805YR7  00000013876YR8  00000013877YR6  00000012796YR9  00000007964YR0  00000008704YR9  00000009948YR1  00000010023YR0  00000010685YR6  00000010833YR2  00000011827YR3  00000011824YR0  00000013057YR5  00000006181YR2  00000006183YR8  00000006184YR6  00000006185YR3  00000006186YR1  00000007951YR7  00000007952YR5  00000008688YR4  00000008689YR2  00000008705YR6  00000008706YR4  00000008707YR2  00000008708YR0  00000008710YR6  00000008711YR4  00000008712YR2  00000008730YR4  00000008731YR2  00000008733YR8  00000009832YR7  00000010227YR7  00000010231YR9  00000010250YR9  00000010253YR3  00000010579YR1  00000010754YR0  00000010755YR7  00000010857YR1  00000010860YR5  00000010862YR1  00000011186YR4  00000011211YR0  00000011212YR8  00000011238YR3  00000011290YR4  00000011343YR1  00000011349YR8  00000011537YR8  00000011539YR4  00000011638YR4  00000011639YR2  00000011649YR1  00000011652YR5  00000011653YR3  00000011685YR5  00000011689YR7  00000011690YR5  00000011828YR1  00000011840YR6  00000011961YR0  00000011963YR6  00000011964YR4  00000011987YR5  00000011991YR7  00000011992YR5  00000012103YR8  00000012117YR8  00000012118YR6  00000012119YR4  00000012141YR8  00000012147YR5  00000012171YR5  00000012366YR1  00000012368YR7  00000012389YR3  00000012390YR1  00000012391YR9  00000012449YR5  00000012551YR8  00000012579YR9  00000012580YR7  00000012581YR5  00000012582YR3  00000012723YR3  00000012729YR0  00000012875YR1  00000012876YR9  00000012879YR3  00000012880YR1  00000012936YR1  00000012937YR9  00000012993YR2  00000013055YR9  00000013092YR2  00000013135YR9  00000013238YR1  00000013240YR7  00000013244YR9  00000013245YR6  00000013249YR8  00000013303YR3  00000013309YR0  00000013347YR0  00000013377YR7  00000013397YR5  00000013445YR2  00000013446YR0  00000013448YR6  00000013452YR8  00000013576YR4  00000013577YR2  00000013578YR0  00000013579YR8  00000013580YR6  00000013582YR2  00000013585YR5  00000013617YR6  00000013618YR4  00000013619YR2  00000013620YR0  00000013622YR6  00000013695YR2  00000013696YR0  00000013697YR8  00000013699YR4  00000013700YR0  00000013701YR8  00000013702YR6  00000013810YR7  00000013914YR7  00000013915YR4  00000013916YR2  00000013935YR2  00000013936YR0  00000013937YR8  00000014058YR2  00000014059YR0  00000014069YR9  00000014104YR4  00000014105YR1  00000014109YR3  00000014110YR1  00000014112YR7  00000014212YR5  00000014213YR3  00000014215YR8  00000014216YR6  00000014217YR4  00000014278YR6  00000014327YR1  00000014329YR7  00000014330YR5  00000014331YR3  00000014332YR1  00000014349YR5  00000014350YR3  00000014351YR1  00000014402YR2  00000014413YR9  00000014416YR2  00000014457YR6  00000014515YR1  00000014519YR3  00000014525YR0  00000014526YR8  00000014554YR0  00000014555YR7  00000014623YR3  00000014654YR8  00000014655YR5  00000014656YR3  00000014657YR1  00000014658YR9  00000014698YR5  00000014699YR3  00000014700YR9  00000014701YR7  00000014702YR5  00000014703YR3  00000014704YR1  00000014705YR8  00000014749YR6  00000014750YR4  00000014751YR2  00000014752YR0  00000014754YR6  00000014800YR7  00000014802YR3  00000014803YR1  00000014804YR9  00000014860YR1  00000014861YR9  00000014862YR7  00000014863YR5  00000014911YR2  00000014957YR5  00000014981YR5  00000014982YR3  00000014983YR1  00000014984YR9  00000014999YR7  00000015000YR3  00000015002YR9  00000015003YR7  00000015041YR7  00000015131YR6  00000015132YR4  00000015133YR2  00000015144YR9  00000015145YR6  00000015226YR4  00000015227YR2  00000015228YR0  00000015229YR8  00000015243YR9  00000015244YR7  00000015302YR3  00000015303YR1  00000015305YR6  00000015309YR8  00000015352YR8  00000015353YR6  00000015363YR5  00000015364YR3  00000015367YR6  00000015409YR6  00000015410YR4  00000015411YR2  00000015412YR0  00000015417YR9  00000015458YR3  00000015459YR1  00000015460YR9  00000015593YR7  00000015633YR1  00000012578YR1  00000012824YR9  00000013801YR6  00000013806YR5  00000014107YR7  00000009947YR3  00000013183YR9  00000006164YR8  00000006509YR4  00000012759YR7  00000013879YR2  00000006202YR6  00000006170YR5  00000008717YR1  00000013625YR9  00000005078YR1  00000005655YR6  00000008690YR0  00000009954YR9  00000006536YR7  00000008693YR4  00000009955YR6  00000010756YR5  00000010828YR2  00000011288YR8  00000011347YR2  00000011497YR5  00000011636YR8  00000011768YR9  00000011825YR7  00000012821YR5  00000013134YR2  00000013181YR3  00000006161YR4  00000006167YR1  00000012291YR1  00000012501YR3  00000012549YR2  00000013311YR6  00000010582YR5  00000011042YR9  00000010832YR4  00000011285YR4  00000011651YR7  00000011966YR9  00000012086YR5  00000012169YR9  00000012288YR7  00000012327YR3  00000012329YR9  00000012506YR2  00000012655YR7  00000012989YR0  00000013060YR9  00000013242YR3  00000013247YR2  00000013248YR0  00000013698YR6  00000014908YR8  00000011038YR7  00000012654YR0  00000014516YR9  00000015304YR9  00000011766YR3  00000013130YR0  00000005073YR2  00000006190YR3  00000006512YR8  00000006520YR1  00000006549YR0  00000010864YR7  00000011839YR8  00000011965YR1  00000012168YR1  00000012293YR7  00000012436YR2  00000012496YR6  00000013395YR9  00000013398YR3  00000006552YR4  00000004672YR2  00000005065YR8  00000005067YR4  00000005069YR0  00000005076YR5  00000005084YR9  00000005645YR7  00000005650YR7  00000005652YR3  00000005653YR1  00000005654YR9  00000005658YR0  00000005660YR6  00000005662YR2  00000006158YR0  00000006159YR8  00000006160YR6  00000006163YR0  00000006165YR5  00000006166YR3  00000006168YR9  00000006178YR8  00000006179YR6  00000006180YR4  00000006182YR0  00000006187YR9  00000006192YR9  00000006193YR7  00000006201YR8  00000006505YR2  00000006506YR0  00000006508YR6  00000006513YR6  00000006517YR7  00000006518YR5  00000006524YR3  00000006526YR8  00000006529YR2  00000006531YR8  00000006537YR5  00000006540YR9  00000006546YR6  00000006548YR2  00000007945YR9  00000007946YR7  00000007948YR3  00000007949YR1  00000007950YR9  00000007953YR3  00000007955YR8  00000007956YR6  00000007957YR4  00000007958YR2  00000007959YR0  00000007960YR8  00000007961YR6  00000007962YR4  00000007963YR2  00000007965YR7  00000007972YR3  00000007973YR1  00000007974YR9  00000008685YR0  00000008686YR8  00000008687YR6  00000008691YR8  00000008696YR7  00000008718YR9  00000008723YR9  00000008725YR4  00000009949YR9  00000009956YR4  00000010021YR4  00000010251YR7  00000010684YR9  00000010687YR2  00000010858YR9  00000010859YR7  00000010861YR3  00000011040YR3  00000011041YR1  00000011289YR6  00000011344YR9  00000011345YR6  00000011346YR4  00000011348YR0  00000011430YR6  00000011672YR3  00000011694YR7  00000011695YR4  00000011764YR8  00000011841YR4  00000011843YR0  00000011988YR3  00000011990YR9  00000011993YR3  00000012082YR4  00000012084YR0  00000012143YR4  00000012148YR3  00000012170YR7  00000012324YR0  00000012325YR7  00000012326YR5  00000012331YR5  00000012653YR2  00000012657YR3  00000012726YR6  00000012730YR8  00000012804YR1  00000012822YR3  00000012939YR5  00000012941YR1  00000013027YR8  00000013028YR6  00000013093YR0  00000013178YR9  00000013182YR1  00000013186YR2  00000013205YR0  00000013250YR6  00000013251YR4  00000013287YR8  00000013344YR7  00000013345YR4  00000013387YR6  00000013644YR0  00000013583YR0  00000014805YR6  00000012292YR9  00000005657YR2  00000006171YR3  00000006173YR9  00000006197YR8  00000007966YR5  00000007967YR3  00000010829YR0  00000010978YR5  00000011039YR5  00000011534YR5  00000011646YR7  00000011691YR3  00000011994YR1  00000012208YR5  00000012392YR7  00000012498YR2  00000012502YR1  00000012505YR4  00000012724YR1  00000013302YR5  00000013447YR8  00000013694YR5  00000013704YR2  00000013803YR2  00000014108YR5  00000014418YR8  00000014454YR3  00000011765YR5  00000006189YR5  00000011286YR2  00000011986YR7  00000012795YR1  00000012817YR3  00000013184YR7  00000013386YR8  00000013396YR7  00000013449YR4  00000013624YR2  00000014062YR4  00000014111YR9  00000014401YR4  00000014403YR0  00000014455YR0  00000014517YR7  00000014912YR0  00000013095YR5  00000015091YR2  00000011767YR1  00000010977YR7  00000011538YR6  00000012226YR7  00000012577YR3  00000012658YR1  00000013094YR8  00000013203YR5  00000012573YR2  00000013381YR9  00000013645YR7  00000012205YR1  00000012938YR7  00000014063YR2  00000011035YR3  00000011498YR3  00000011962YR8  00000011989YR1  00000013243YR1  00000013343YR9  00000006539YR1  00000006551YR6  00000008719YR7  00000008721YR3  00000010254YR1  00000011037YR9  00000012754YR8  00000014060YR8  00000014071YR5  00000015092YR0  00000006156YR4  00000010225YR1  00000012547YR6  00000012819YR9  00000005077YR3  00000006169YR7  00000008694YR2  00000008713YR0  00000008714YR8  00000010226YR9  00000011432YR2  00000012210YR1  00000004773YR8  00000005082YR3  00000006196YR0  00000006521YR9  00000006545YR8  00000006550YR8  00000010230YR1  00000012823YR1  00000013304YR1  00000013307YR4  00000013376YR9  00000010229YR3  00000014211YR7  00000012083YR2  00000013179YR7  00000012328YR1  00000012990YR8  00000013450YR2  00000014753YR8  00000015001YR1  00000006527YR6  00000006547YR4  00000011532YR9  00000013382YR7  00000013623YR4  00000013627YR5  00000013628YR3  00000013703YR4  00000013807YR3  00000014061YR6  00000014064YR0  00000010252YR5  00000012797YR7  00000005072YR4  00000005074YR0  00000006203YR4  00000006510YR2  00000006511YR0  00000006523YR5  00000006530YR0  00000006533YR4  00000006534YR2  00000006541YR7  00000006542YR5  00000006543YR3  00000006553YR2  00000006554YR0  00000010939YR7  00000010976YR9  00000012552YR6  00000012553YR4  00000012574YR0  00000012793YR6  00000012794YR4  00000012878YR5  00000012996YR5  00000013058YR3  00000013059YR1  00000013204YR3  00000013378YR5  00000013629YR1  00000013757YR0  00000014328YR9  00000014417YR0  00000014456YR8  00000011838YR0  00000006199YR4  00000010115YR4  00000011536YR0  00000012290YR3  00000014518YR5  00000015533YR3  00000015534YR1  00000015535YR8  00000015536YR6  00000015576YR2  00000015577YR0  00000015578YR8  00000015579YR6  00000015632YR3  00000015644YR8  00000015687YR7  00000015688YR5  00000015693YR5  00000015694YR3  00000015725YR5  00000015746YR1  00000015868YR3  00000006175YR4  00000006195YR2  00000008700YR7  00000008722YR1  00000008724YR7  00000010581YR7  00000011496YR7  00000011668YR1  00000012101YR2  00000012145YR9  00000012146YR7  00000012149YR1  00000012792YR8  00000013308YR2  00000014065YR7  00000014458YR4  00000014622YR5  00000014621YR7",
      "more_code_info": ""
    }
  ]
}