  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82491",
      "recalling_firm": "Carestream Health, Inc.",
      "address_1": "1049 W Ridge Rd",
      "address_2": "N/A",
      "postal_code": "14615-2731",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) to states of: Arizona, California, New York, Alabama, Georgia, Georgia, Illinois, Minnesota, New York, Oregon, Pennsylvania; and countries of: IRAN, ITALY, JERUSALEM, KOREA, PHILIPPINES, PORTUGAL,SAUDI ARABIA, SINGAPORE, SWITZERLAND  THAILAND, and UNITED KINGDOM.",
      "recall_number": "Z-1207-2019",
      "product_description": "OnSight 3D Extremity System- X-Ray, Tomography Computed System  Catalog Numbers:  1743566 (w/ Attached Console)  1743574 (w/ Remote Console)    The device is intended to be used for x-ray computed tomography and projection x-ray imaging of upper and lower extremities of adult patients and pediatric patients aged 12 and over.",
      "product_quantity": "37 units",
      "reason_for_recall": "When the user performs the re-assignment of a parent / companion pair, the parent volume is transferred to the new patient but the companion volume will remain in the original patient s exam.",
      "recall_initiation_date": "20190319",
      "center_classification_date": "20190423",
      "termination_date": "20200427",
      "report_date": "20190501",
      "code_info": "US Serial Numbers:  58008585  58007506  58000020  58008493  58000015  58007502  58007504  58008494  58007503  58008018  58007505  58008500  58008011  58008014    OUS:  10012  20005  20012  20016  20007  20017  10018  10004  10007  20004  20008  10023  10009  20019  20013  10002  10003  10014  10010  20006  20022  20003  20010"
    }
  ]
}